![nanoString nCounter Pro Скачать руководство пользователя страница 74](http://html1.mh-extra.com/html/nanostring/ncounter-pro/ncounter-pro_user-manual_3235005074.webp)
nCounter Pro Analysis System User Manual
SEV-00184-01
FOR RESEARCH USE ONLY. Not for use in diagnostic procedures.
74
Email accounts can be deleted from the Prep
Station as well. To delete an email address, press
the email accounts button from the ‘Email
configuration’ screen (Figure 110).
The ‘Email Accounts’ screen will appear
(Figure 111).
To delete an email account from the Prep Station,
select that email address and press delete.
System Hardware
To access the System Hardware Menu, press
hardware on the Maintenance Menu. The System
Hardware Menu (Figure 112) enables the user to:
•
home robot: Homes all the motors should
the robot enter an undesired state.
•
align electrodes: Guides the user in how to
align electrodes if they become bent and no
longer fit into the cartridge.
•
lubricate O-rings: Alerts the user when the
O-rings need to be lubricated and guides
the user in the process.
•
calibration check: NanoString recommends that customers do not run the calibration check
program. If the Prep Station has been idle for more than 1 month, it is recommended to
perform a
Figure 111: The ‘Email Accounts’ screen
Figure 112: The System Hardware Menu.