nCounter Pro Analysis System User Manual
FOR RESEARCH USE ONLY. Not for use in diagnostic procedures.
63
Pausing a Run
From the ‘Counting Cartridge’ screen (Figure 92), runs can be paused to perform any of several
actions:
•
add cartridge: A cartridge can be added if, for example, the device is in use, but the Prep
Station has recently finished processing samples.
•
abort cartridge: A single cartridge can be aborted; cartridges may be re-imaged.
•
abort run: The entire run can be aborted; cartridges may be re-imaged, and data from
cartridges that have already been imaged will be saved.
•
upload files: New CDFs or RLFs can be uploaded if, for example, incorrect files were originally
assigned when the run was initiated.
NOTE:
The pause button is present while the run is still in progress. Once the run is
complete, the pause button will be replaced by the finished button.
Add a New Cartridge
1.
On the ‘Counting Cartridge ID’ screen, press
pause.
The ‘Imaging Paused’ screen will appear after
the imaging device halts (Figure 94).
2.
Press add cartridge.
3.
After completing the ‘add cartridge’ workflow
(identical to Steps 2–7 for defining a new
cartridge in the ‘start counting’ workflow),
press resume run to continue processing.
Figure 94: The ‘Imaging Paused’ screen.