10. Option to set/calculate a dilution factor for manual diluted samples.
11. Apply the blank ddH
2
0 or buffer to the illuminated sample window for the
reference measurement and select blank to initiate the reading.
Note:
The illumination of the sample window can be switched off in the
preferences.
12. Use a lint-free laboratory wipe to clean both the sample window and
mirror prior to applying the next sample.
Note:
It can be helpful to apply the blank a second time and read it as a
sample to ensure a proper blank.
13. Apply sample to the sample window and press the sample button to
initiate the measurement.