26
Introduction
Trouble shooting
Trouble
shooting
Problem
Description
Solution
ADAM-MC2 does
not power up
•
No power from outlet
•
Bad power cord.
•
Check power source.
•
Replace.
Inaccurate result
•
Cell number may be out of
range.
•
AccuStain Solution has
expired.
•
Too high clumped cells.
•
Adjust the number of cells
between 5X10 ~ 4X10
cells/mL (refer to page 12).
•
Discard AccuStain that
have expired. Purchase the
AccuStain(ADR-1000).
•
Try again after vortexing
the cells.
When error
message is shown
(For information on
each error message,
see page 17)
• When frames with errors
are over 50% of total
counting frame.
(Error message: E)
•
Check the suspension of
cells if all cells are fully
dissociated into single cells.
•
If contaminants except
cells are found, prepare
sample again.
• High concentration of cells
(Error message: H)
• Over detection range
(Error message: O)
• Check if concentration of
cell is too high.
• Dilute the sample and
count again.
•
Low concentration of cells
(Error message: L)
•
Under detection range
(Error message: U)
•
Check if concentration of
cell is too low.
•
Use concentrated sample
and count again.
4