nitrogen gas generator
ECOGEN
2
4.2 remote start/stop control
To gain access to the remote start/stop feature:
•
Remove the two M6 screws from each side of the enclosure.
•
Open the front cover and locate the flying lead on the bottom of the control plate and remove the insulating sheath.
•
There are six wires;
1. Red Wire
- Remote Stop
2. Black Wire - Remote Start/Stop
3. Blue Wire - Alarm Output (Zero volt contact)
4. White Wire - Alarm Input (Zero volt contact)
5. Green Wire - 24V DC Output
6. Yellow Wire - 24V DC Output
•
To set up the Remote Start/Stop control, remove/break the connection between the Black wires and green wires connect
externally to a remote switch or relay.
•
A 24V DC Output must be connected to the Black wire to enable the dryer to operate, if the connection is broken or if there is no
voltage the generator will switch off and revert to shutdown mode, displaying
“REMOTE STOP ACTIVE”
on the controller display.
IMPORTANT:
Under no circumstances should an external voltage or current be applied to any of these wires, as
damage to the control system will occur, negating the warranty.