EN | BM-03 Baby breathing monitor
MXA51006
(85956141BM03SJ)
6.1
LIFE SPAN OF THE SENSOR PAD
7.
POWER SUPPLY AND REPLACING THE BATTERIES
The BM-03 Baby breathing monitor tracks the number of hours the monitor has been
in operation as an auxiliary data - if the sensor pad has probably exceeded the specifi ed
period of use, 3x warning beeps sound during each automatic functionality test.
However, the date it is put into service is decisive for determining the life span.
The sensor in the pad may wear out over time, so it may not correctly sense your baby’s
movements and breathing. In this case, the monitor may give false alarms that will cause
you unnecessary worries.
False alarms
may not endanger your baby’s life or health, but
you will hear the alarm so often that you prefer to turn the monitor off or reduce your
alertness, which could endanger your child.
The device does not require any special maintenance other than replacing the batteries
and cleaning. We simply recommend that you check the sensor pad in the cot from time
to time to ensure moisture has not condensed where the pad touches the mattress. It is
advisable to turn the mattress 180° in the cot once in a while, or turn it upside down or let
it air, etc. Only use a slightly damp cloth for cleaning. Avoid wet wipes or other materials
that may have loose fi bres that can clog the holes in the control unit.
It is not necessary to sterilize the breathing monitor when in medical facilities. The device
does not require sterilization or disinfection between use by different patients. The device
can be treated
with commonly available disinfectants
at a concentration in line with
the manufacturer’s recommendations for the given level of disinfection. Keep in mind
that moisture entering the battery compartment or the control unit during cleaning can
damage the device. Only use a cloth dampened with disinfectant solution for disinfection
and only wipe the device lightly.
The frequency of cleaning is not specifi ed by the manufacturer and the frequency of
cleaning does not affect the product’s life span. As concerns further maintenance, the
device does not have a measuring function to be calibrated.
The life span of the sensor pad is set at 2 years, after which the pad must be
replaced.
During maintenance, protect the pad, supply cable and connectors from
mechanical damage and moisture entering.