51
EN
MXA51006
(85956141BM03SJ)
The alarm can be
turned off
by pressing the mechanical on/off button. The monitor
operates in active and standby mode. In
active mode
, the monitor detects the baby’s
breathing movements and additional functions are activated. In
standby mode
, the
monitor is switched to power saving mode and the unit evaluates the signals from the
pad to be able to draw attention to a baby being placed there.
The transition to standby mode is indicated by an optical and acoustic signal.
Permanently jamed mechanical button indicates a malfunction. If this fault is detected
during the automatic functionality test after the standby mode transition, the function of
the evaluation unit will not be triggered. If a mechanical button fault is detected in active
mode, a critical failure will be reported visually and acoustically. Release the mechanical
button to remove the fault. Press the button once more to turn off the malfunction warning.
Automatic functionality test
When switching from standby to active mode or when inserting the batteries, the
device automatically tests its functionality. The test checks the batteries, the correct type
of pad is connected and its age, the optical and acoustic signal, the notifi cation for the
on/off switch being in the incorrect state for the baby’s current absence or presence.
Functionality test results:
A.
Flashing of all indicator lights, a short beep and 10x fl ashes of the indicator light
with the power-on symbol = all checks have been carried out correctly and the
device is fully functional.
B.
Flashing lights and repeated warning beeps = error diagnosed but it does not
prevent the monitor from being used and maintaining its functionality.
C. One or two warning beeps, orange fl ashing of the indicator light around the
mechanical button designed to turn the monitor on/off and the device does
not turn on = critical error detected (2 beeps critical battery condition, 1 beep
unconnected sensor pad), device cannot be used. Correct the error (insert new
batteries or connect the sensor pad), then you can use the device again. If the error
could not be removed, send the monitor to the service center.
2.2.1
DETECTING WHETHER THE BABY HAS BEEN PLACED ON THE SENSOR PAD OR
TAKEN OFF IT (SWITCH GUARD)
When in standby mode the BM-03 Baby breathing monitor continuously evaluates
signals from the pad and is thus the only device on the market today to point out that
Type of warning
Acoustic indication
None – the device is fully functional
1x beep
The Switch Guard function is deactivated
2x beep
Connected a sensor pad or control unit with possibly exceeded
life span
3x beep
2.2
SUPPLEMENTARY FUNCTIONS