CAUTION
• Make sure to turn the switching valve until it
stops.
• Before initiating treatment, assure the water is
spraying correctly.
• If water is not being supplied correctly, recheck
the switching valve.
• Be careful of overheating when using some types
of carbide bur, as the coolant water (in “water
jet” mode) cannot reach the bur tip because of
the length of the tip. If the water does not reach
the tip, select “water spray” mode.
• For the water jet shape, set the water
consumption to 50 mL/min or more to provide
sufficient cooling.
8
Post-use Maintenance
After each patient maintain the product as follows.
WARNING
• To ensure cleaning and sterilization efficacy, use
only the following procedures for reprocessing.
• Follow any additional local directives, standards,
and guidelines for cleaning and sterilization.
• Clean and lubricate the handpiece immediately
(within 1 hour) after each treatment to remove
residue. Failure to properly maintain the
handpiece may cause infection, product failure,
overheating leading to burn injuries.