13
Failure phenomenon
Fault analysis
Troubleshooting
Open machine without
display
Power supply is not connected
Check if the power outlet with voltage
The power plug is not plugged
in
Check the reliability of the contact
between the power plug and the socket
The power switch is not open
Turn on the power switch on the right side
of the instrument
The damage of the fuse on the
box
Replacement of a power fuse with the
same specification
Rolling plate does not rotate
Door switch does not work
Check whether the door is closed or not.
The bottom of the rolling plate
is stuck with foreign objects
Remove foreign body
Belt damage
Notify distributor to replace the belt
Control circuit fault
Notify distributor to repair
Instability of rolling plate
The instrument is not placed
smoothly
Adjust the foot of the instrument to make
the instrument smooth
Foreign body barrier at the
bottom of the plate
Remove foreign body
Control circuit fault
Notify distributor to repair
The measured temperature is
higher than the set
temperature or the instrument
enters the high temperature
alarm state
The door is not closed well
Close the door well
The instrument has not yet
entered the state of constant
temperature
Wait for a moment to observe again
Damage of circulating fan
Notify distributor to replace the fan
The noise of the incubator is
too loud
The incubator is not placed
smoothly
Adjust the foot to make the incubator
smooth
Flask clamp fixing screw loose
Remove the rocker with the tool fasten
screw
Shake plate loosening
Tighten the fixing screws at the four
corners of the rocking plate
Bump on the bottom of the
plate with foreign objects
Remove foreign body
Mechanical failure
Notify distributor to repair
6
.