11
6. Operation control: Press the “Start/Stop” button, the identifier “RUN” at the lower right of
the interface will light up and start to turn; press the “Run/Stop” button again, the identifier
“STOP” at the bottom right of the interface will light and stop rotating.
7. Gating: During operation, open the door, the interface identifier “OPEN” lights up, the
rotation stops, the door is closed, the identifier “OPEN” goes out, and the rotation continues.
8. Lighting: When “Lt=0”, click the “Lighting” button to turn the light on and off. The
illumination off delay setting is set by referring to the description of “Temperature---Internal
Parameter Table (1)”.
5. Function setting description
1. Timing function
When the total timing time is set to “0”, it means that there is no timing function, the device
runs continuously; when the total timing time is not “0”, the timing is started according to the
timing mode, “m” (time in minutes) or “ h” (timed by hour) flashes, the motor stops when the
timer ends, the time display area displays
“
End
”
, the buzzer sounds, and the “start/stop”
button is clicked to restart the operation.
Notes: 1) If the set value of the timing time is changed during the running, the controller
restarts the timing according to the new set time.
2) The timing mode can be selected by modifying the "ndt" value in the "Time--Internal
Parameter Table".
3) The timing mode can be selected by modifying the "Hn" value in the "Time - Internal
Parameter Table".
2. Power failure memory function
You can select whether there is a power-down memory function by modifying the value of
“
rES
”
in
“
Temperature-Internal Parameter Table (2)”.
1) No power-down memory function (rES=0): the controller is powered off after power
failure, and the speed is stopped.
2) There is power-down memory function (rES=1): the controller is powered off after
power-off, and the speed is in the running state at the last power-off.
6. Alarm and shutdown
1.
When the power module failure (Er-1), Er-2, Holzer error (Er-3), busbar undervoltage
(Er-4), busbar overvoltage (Er-4), busbar overvoltage (Er-5), communication fault (Er-6), the
identifier "ALM2" is lit, the controller is automatically stopped and the buzzer calls. Please
turn off the power when the fault occurs, carefully check the wiring of the motor, and make
sure it is correct again.
2. When the over temperature alarm occurs, the buzzer sounds and the "ALM1" identifier