10
ndt
Timing mode
selection
0: Start timing after running, running time is up, only stop
speed, non-stop temperature
1: Start timing after running, run time arrives, stop speed
and temperature simultaneously
2: The temperature starts to count after reaching the set
value. The running time is up, only stop speed, non-stop
temperature.
3: When the temperature reaches the set value, the timing
starts, the running time arrives, and the speed and
temperature are stopped at the same time.
When the timing starts,
“
m
”
or
“
h
”
starts to flash, and
after the time is over, the time display area displays
“
End
”
.
Note: This parameter is prohibited from being modified
during operation.
(0
~
3)
1
Hn-
Total timing
Timing mode
0: Timing is measured in minutes
1: Timing is measured in hours
Note: This parameter is prohibited from being modified
during operation.
(0
~
1) 0
rT-
Total timing
Time correction
Correct the total timing timing error,
Correction value = [run time (seconds) - actual time
(seconds)] * 10 / actual time (minutes)
(-999
~
999) 0
4. Buttons and operation instructions
1. Power on: Turn on the power, the controller is powered on, the temperature display window
displays
“
HY-2
”
, the speed display window displays
“
8888
”
, the time display window
displays
“
8888
”
, all the identifiers light up, and enters after about 3 seconds. Normal display
status.
2. Temperature: Press the
“
Temperature
”
button, the number on the right side of the
temperature display area flashes, use the
“
◀
”
shift button,
“▼”
decrease button,
“▲”
increase button to set the desired temperature, then press the
“
Temperature
”
button exits the
temperature setting state, and the set temperature value is automatically saved, and the
instrument operates according to this temperature value.
3. Speed: Press the
“
Speed
”
button, the number on the right side of the speed display area
flashes, use
“
◀
”
shift button,
“▼”
decrease button,
“▲”
increase button to set the
required speed, press again the
“
Speed
”
button exits the speed setting state, and the set speed
value is automatically saved, and the instrument runs at this speed value.
4. Time: When the total timing is set to
“
0
”
, it means that there is no timing function and the
equipment is running continuously. The timing function setting is set by referring to the
description in item 8 above.
5. Sterilization: When
“
St=0
”
, long press the
“
sterilization
”
button for 3 seconds, the
identifier
“
STERILIZE
”
under the interface lights up, the UV lamp is turned on, the device
is in sterilization operation; and then press the
“
sterilization
”
button. In 3 seconds, the
identifier "STERILIZE" is extinguished and sterilization is stopped. The sterilization lamp off
delay setting is set by referring to the description of "Temperature - Internal Parameter Table
(1)".