Set the required temperature using the numeric keys
and press ENTER.
The tower will heat to SET temperature.
Wait until the required SET temperature is equal to
the READ temperature.
When the SET and the READ temperature have been
equal for about 07 minutes, a beep signal will heard
and a tick mark will appear in the display. This
indicates that you have achieved the stable
thermowell temperature. Now you can take the
readings.
Connect the RTD probes to the corresponding
terminals in the front plate of the instrument.
Display shows the bath temperature as the INPUT
parameter which corresponds to the READ
temperature.
Note: If you select mvL or mvH while using T/C, add
the voltage corresponding to the ambient
temperature with the INPUT voltage which is shown
in the display.