7.
OPERATING INSTRUCTION
Precheck
Switch ON the instrument.
Wait until the self-diagnostic routines are finished.
Set value (default) should be zero.
Read value should show the ambient temperature.
Set any temperature in the display.
If the up arrow shows the display, then we can make
sure that the instrument is in working Condition.
Preparation of the Calibrator for Calibration
Insertion tube diameter is selected according to the
diameter of the sensor to be calibrated.
The fan should begin quietly blowing air through the
instrument after the illumination of controller
display.
Sensor is inserted in the insertion tube
Display shows SET as 0˚C and READ as Room
Temperature.
Select the required parameters by pressing SELECT
option in the keypad.
It shows the parameters such as T/C, RTD, mA, mvL,
mvH, volts, mA, ohms, kohms, T/C and RTD. Select
the sensor and its type which is to be calibrated.
Choose the read-out scale (°C/°F) you require by
activating the key