![multicomp pro MP720105 Скачать руководство пользователя страница 85](http://html1.mh-extra.com/html/multicomp-pro/mp720105/mp720105_manual_1854300085.webp)
4.Advanced User Guidebook
Each parameter name is a case-sensitive string of 4 bytes. The parameter value is at
least 4 bytes.
1.
Format description of the file header:
1)
HEAD
Parameter name Meaning
Value
Comment
HEAD
Header size
4 bytes int
2) TYPE
Parameter name Meaning
Value
Comment
TYPE
Model
12 bytes char
3) BYTE
Parameter name Meaning
Value
Comment
BYTE
Data length in bit
4 bytes int
4) SIZE
Parameter name Meaning
Value
Comment
SIZE
File size
4 bytes int
Used to check the file
integrity
5) VOLT
Parameter name Meaning
Value
Comment
VOLT
Voltage division,
divided by 400 is
ADC resolution.
(When the source is
CH1&CH2, it is CH1
voltage division.)
4 bytes float
The value indicates voltage
(the unit is mV), such as 200
mV.
6) SAMP
Parameter name Meaning
Value
Comment
SAMP
Sample rate
4 bytes float
The unit is Sa/s.
7) ADCB
Parameter name Meaning
Value
Comment
ADCB
ADC bit, ADC
resolution
4 bytes int
8-bit or 12-bit
8)
CHAN
Parameter name Meaning
Value
Comment
CHAN
Number of channels 4 bytes int
1 or 2
9)
VOL2
Parameter name Meaning
Value
Comment
VOL2
Voltage division,
divided by 400 is
ADC resolution.
(When the source is
CH1&CH2, it is CH2
voltage division.)
4 bytes float
The value indicates voltage
(the unit is mV), such as 200
mV.
80
Содержание MP720105
Страница 1: ...i Digital Storage Oscilloscope Model MP720105 MP72016 MP72107...
Страница 40: ...4 Advanced User Guidebook Figure 4 2 Rectangle window Figure 4 3 Blackman window 35...
Страница 41: ...4 Advanced User Guidebook Figure 4 4 Hanning window Figure 4 5 Kaiser window 36...
Страница 68: ...4 Advanced User Guidebook Sine x x interpolation Linear interpolation 63...