![multicomp pro MP720105 Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/multicomp-pro/mp720105/mp720105_manual_1854300008.webp)
2.Safety Terms and Symbols
The diagram of the oscilloscope ground wire connection:
Ground Clip
Signal Input
Electrical Outlet
Probe Oscilloscope
Power Cord
The diagram of the ground wire connection when the battery-powered oscilloscope
is connected to the AC-powered PC through the ports:
Ground Clip
Signal Input
Oscilloscope
(Battery-power)
PC
Electrical Outlet
Probe
USB/VGA/COM/
LAN Cable
It is not
permitted
to measure AC power when the oscilloscope is AC powered,
or when the battery-powered oscilloscope is connected to the AC-powered
PC through the ports.
Warning:
To avoid fire or electrical shock, when the oscilloscope input
signal connected is more than 42V peak (30Vrms) or on circuits of
more than 4800VA, please take note of
the following
items:
Only use accessory insulated voltage probes and test lead.
Check the accessories such as probe before use and replace it
if there are any damages.
Remove probes, test leads and other accessories immediately after
use.
Remove USB cable which connects oscilloscope and computer.
Do not apply input voltages above the rating of the
instrument because the probe tip voltage will directly
transmit to the oscilloscope. Use with caution when the probe is
set as 1:1.
Do not use exposed metal BNC or banana plug connectors.
Do not insert metal objects into connectors.
3
Содержание MP720105
Страница 1: ...i Digital Storage Oscilloscope Model MP720105 MP72016 MP72107...
Страница 40: ...4 Advanced User Guidebook Figure 4 2 Rectangle window Figure 4 3 Blackman window 35...
Страница 41: ...4 Advanced User Guidebook Figure 4 4 Hanning window Figure 4 5 Kaiser window 36...
Страница 68: ...4 Advanced User Guidebook Sine x x interpolation Linear interpolation 63...