![multicomp pro MP720105 Скачать руководство пользователя страница 65](http://html1.mh-extra.com/html/multicomp-pro/mp720105/mp720105_manual_1854300065.webp)
4.Advanced User Guidebook
4. CAN Decoding
To decode CAN signal:
(1)
Connect the CAN signal to the Signal Input Channel of the oscilloscope.
(2)
Adjust to the proper time base and voltage division.
(3)
In trigger menu, select Bus trigger, and select bus type as CAN, set parameters
based on the characteristics of the signal,
trigger the signal correctly and obtain
stable display. Refer to "
(4)
Push the
Decode
button on the front panel. Select bus type as CAN. set
parameters based on the characteristics of the signal. When the parameters are set
correctly, the information carried by the signal will be displayed.
Tip:
If there are repetitive menu items in both trigger menu and decoding menu, you
can set anyone of them, the other will be changed synchronously.
Decoded information interpretation:
Information
Abbreviation
Background
Identifier
I, ID, or do not display
Green
Overload Frame
OF
Green
Error Frame
EF
Green
Data Length code
L, DLC, or do not display
Blue
Data
D, Data, or do not display
Black
Cyclic Redundancy Check
C, CRC, or do not display
Valid: Purple
Error: Red
Note
:
Use the
Trigger Level
knob to adjust the thresholds of bus trigger and bus
decoding.
When the ACK (ACKnowledge Character) of Data Frame or Remote Frame is
not met, two red error marks will be
displayed in the corresponding position in
the waveform.
Error Frame, Remote Frame, and Overload Frame will be identified in the "Data"
column in the event table (Data Frame will not be identified).
60
Содержание MP720105
Страница 1: ...i Digital Storage Oscilloscope Model MP720105 MP72016 MP72107...
Страница 40: ...4 Advanced User Guidebook Figure 4 2 Rectangle window Figure 4 3 Blackman window 35...
Страница 41: ...4 Advanced User Guidebook Figure 4 4 Hanning window Figure 4 5 Kaiser window 36...
Страница 68: ...4 Advanced User Guidebook Sine x x interpolation Linear interpolation 63...