Temposonics
®
R-Series
V
EtherNet/IP™
Operation Manual
I 15 I
NOTICE
Use magnets of the same type (e.g. 2 ring magnets ) for
multi-position measurement.
The minimum allowed distance between magnets (i.e. front face of
one to the front face of the next one) is 75 mm (3 in.). for multi-
position measurement.*
*/
Contact MTS Sensors if you need a magnet distance, which is smaller than 75 mm (3 in.).
Controlling design dimensions are in millimeters and measurements in ( ) are in inches
Multi-position measurement
The minimum distance between the magnets is 75 mm (3 in.).
RP5 with U-magnet
≥ 75 (≥ 3)
RP5 with magnet sliders
≥ 75 (≥ 3)
RP5 with block magnets
≥ 75 (≥ 3)
RH5 with ring magnets / U-magnets
≥ 75 (≥ 3)
RH5 with block magnets
≥ 75 (≥ 3)
Fig. 19: Minimum distance for multi-position measurement