CMS-50FF 

 

Instructions for CMS-50FF Pulse Oximeter

 

Dear users, thank you very much for purchasing the Pulse Oximeter.     

This Manual is written and compiled in accordance with the

 

council directive  MDD93/42/EEC for 

medical  devices  and  harmonized  standards.  In  case  of  modifications  and  software  upgrades,  the 

information contained in this document is subject to change without notice. 

The Manual describes, in accordancKe with the  Pulse Oximeter’s features and requirements, main 

structure, functions, specifications, correct methods for transportation, installation, usage, operation, 

repair,  maintenance and  storage, etc.  as  well  as  the safety  procedures  to protect both  the user  and 

equipment. Refer to the respective chapters for details. 

Please read the User Manual carefully before using this product. The User Manual which describes 

the operating procedures should be followed strictly. Failure to follow  the User Manual may cause 

measuring abnormality, equipment damage and human injury. The manufacturer is NOT responsible 

for the safety, reliability and performance issues and any monitoring abnormality, human injury and 

equipment  damage  due  to  users'  negligence  of  the  operation  instructions.  The  manufacturer’s 

warranty service does not cover such faults.     

Owing  to  the  forthcoming  renovation,  the  specific  products  you  received  may  not  be  totally  in 

accordance with the description of this User Manual. We would sincerely regret for that.                                                                                                 

 

This product is medical device, which can be used repeatedly. 

 

WARNING: 

 

Uncomfortable or painful feeling may appear if using the device ceaselessly, especially for 

the  microcirculation  barrier  patients.  It  is  recommended  that  the  sensor  should  not  be 

applied to the same finger for over 2 hours. 

 

For the special patients, there should be a more prudent inspecting in the placing process. 

The device cannot be clipped on the edema and tender tissue. 

 

The light (the infrared is invisible) emitted from the device is harmful to the eyes, so the 

user and the maintenance man should not stare at the light. 

 

Testee cannot use enamel or other makeup. 

 

Testee’s fingernail cannot be too long. 

 

Please refer to the correlative literature about the clinical restrictions and caution. 

 

This device is not intended for treatment.

 

The User Manual is published by our company. All rights reserved. 

Содержание CMS-50FF

Страница 1: ...ty and performance issues and any monitoring abnormality human injury and equipment damage due to users negligence of the operation instructions The manufacturer s warranty service does not cover such...

Страница 2: ...cations 4 4 1Mainperformance 4 4 2MainParameters 4 5Installation 5 5 1Viewofthefrontpanel 5 5 2Wristbandinstallationandprobeconnection 5 5 3Accessories 6 6OperatingGuide 6 6 1Applicationmethod 6 6 2At...

Страница 3: ...attery plastic bags foams and paper boxes should follow the local laws and regulations Please check the packing before use to make sure the device and accessories are totally in accordance with the pa...

Страница 4: ...d default setting for good This function could be turned on through menu operation if you need Please check the chapter 6 1 as a reference The device may not work for all patients If you are unable to...

Страница 5: ...95 c Atmospheric pressure 500hPa 1060hPa Operating Environment a Temperature 10 40 b Relative Humidity 30 75 c Atmospheric pressure 700hPa 1060hPa 3 Principle Principle of the Oximeter is as follows A...

Страница 6: ...rs A Measurement of SpO2 Measuring range 0 100 Accuracy When the SpO2 measuring range is 70 100 the permission of absolute error is 2 Below 70 unspecified B Measurement of pulse rate Measuring range 3...

Страница 7: ...ication light When the device is in the state of battering charging the indication light appears to be orange and when the battery status is full the light turns to green 5 2 Wristband installation an...

Страница 8: ...by our company never replace it with the similar ones by other manufacturers b Put the finger into the probe c Turn on the device by long pressing the button on the panel d Do not shake the finger and...

Страница 9: ...order to enter the clock interface and the device would return to the measuring interface by Press the button for about 10 seconds again D Pause alarm a Alarm including the alarm of measure data s goi...

Страница 10: ...LM HI SpO2 low limit SpO2 ALM LO Pulse rate high limit PR ALM HI Pulse rate low limit PR ALM LO Press the button and hold to adjust the selected limit to the desired higher value and release the butto...

Страница 11: ...lock Setting Menu a When entering the clock setting menu the menu choice bar would be on the item of set time and the state would always be no whenever it enters the clock setting menu on the purpose...

Страница 12: ...Figure 8 SpO2 program Figure 9 SpO2 Review program If the users choose to turn on the synchronizing display function on computer it would probably take several seconds for the data to appear on the co...

Страница 13: ...d to your computer c In the state of storing it is not applicable for the users to upload the stored date to computer d When the upload of stored data is finished the menu choice bar will move to Exit...

Страница 14: ...digo blue or carbon monoxide hemoglobin COHb or methionine Me Hb or thiosalicylic hemoglobin and some with icterus problem the SpO2 determination by this monitor may be inaccurate C The drugs like dop...

Страница 15: ...The batteries are drained or almost drained 2 The device s malfunction 1 Please recharge the battery 2 Please contact the local service center The display is off suddenly 1 The device s malfunction 2...

Страница 16: ...Function Specification Information Display Mode The Pulse Oxygen Saturation SpO2 2 digit digital OLED display Pulse Rate PR 3 digit digital OLED display Pulse Intensity bar graph bar graph OLED displ...

Страница 17: ...ent Voltage 3 7 rechargeable lithium battery 1 Battery working life Charge and discharge no less than 500 times Power Adapter Input Voltage 100 to 240 VAC 50 60 Hz Output voltage 5 VDC Output current...