HYDROCAL 1005 Manual for Installation and Operation
Page 27/111
4.3.3.
Leak detection
During the leak detection cycle the pressure stability of the complete system is checked
Step
No.
Status of the components
[1]
Capacitive Sensor F0
“not active”
[2]
Capacitive Sensor FI
“not active”
[3]
Capacitive Sensor FII
“not active”
[4]
Valve V6
“closed”
[5]
Flow meter
“not active”
[6]
Valve V1
“draining / filling”
[7]
Valve V3
“closed”
[8]
Compressor K1
“not active”
[9]
Valve V4
”closed”
[10]
Valve V2
“open”
[11]
Pump P1
“not active”
[12]
Valve V5
“open”
4.3.3.1. Safety function and error observation for leak detection cycle
Observed Components
Detectable Errors
Filling Sensor 0, 1, 2
Operating in normal parameters
Measurement chamber
Measurement chamber offline
Pressure Sensor
Leakage
Flow meter
V6 closed