Apply a defined gas concentration on the sensor element with the help of the gas generator and check the 
measurement signal at the analog output or at the relay outputs.  

With this 

test, the complete function chain “Sensor Element > Sensor Cartridge> Local Bus> 

PCB> Field Bus> 

GC Controller” i

s tested. 

 

6

 

Inspection and Service 

Inspection, service and calibration of the sensor should be done by trained technicians at regular intervals. 
We therefore recommend concluding a service contract with MSR or one of their authorized partners. 

According  to  EN  45544-4,  inspection  and  service  have  to  be  executed  at  regular  intervals.  The  maximum 
intervals  have  to  be  determined  and  respected  by  the  person  responsible  for  the  gas  warning  system 
according  to  the  legal  requirements.  MSR-E  recommends  employing  the  common  inspection  and  service 
intervals  as  specified  in  the  general  regulations  of  the  gas  measuring  technique.  We  recommend  an 
inspection interval of 3 months. The recommended calibration intervals depend on the sensor element and 
can be read from the tabl

e “Calibration”. If there are different

 intervals, always observe the shortest one. 

Inspections and services must be documented. The date for the next maintenance has to be affixed to the 
sensor. 

6.1

 

Inspection 

Gas sensors should be controlled regularly by a competent person according to EN 45544-4. The following 
has to be checked in particular: 

 

Maintenance/ calibration interval not exceeded. 

 

Visual inspection of the sensor including cable for damage etc. 

 

Remove dust deposits, especially at the gas inlet.  

 

The filter at the gas inlet has to be replaced if extremely dirty. 

6.2

 

Service and Calibration 

When performing the maintenance, you have to do the calibration and the functional test, see chapter 5, in 
addition to the inspection. 

A fixed calibration interval is stored for each sensor type. 

SC2 Sensor heads: 

If this interval is exceeded, a digital maintenance message is generated and forwarded. 

Performing the calibration automatically deletes this message. 

MC2 Sensor heads: 

If  this  interval  is  exceeded,  the  current  output  of  the  MC2  goes  to  a  fixed  19  mA  signal.  For  new  devices 
(factory calibration), the interval may be exceeded by a factor of 1.5. 

After voltage recovery, there is a defined interruption of this message. This can be of use in order to employ 
the device without the maintenance message for a few days (adjustable in system parameters -> error time, 
value range> 1) until the calibration is repeated. 

Performing the calibration automatically deletes this message. 

Содержание PolyGard 2

Страница 1: ...t to change without notice MSR Electronic GmbH Buergermeister Schoenbauer Str 13 D 94060 Pocking www msr electronic de PolyGard 2 Gard 2 Sensors for Toxic Gases Sensor Cartridge for Toxic Gases and Ox...

Страница 2: ...n SB2 MSB2 MSC2 WSB2 5 3 2 Terminal Connection MC2 5 4 Commissioning 5 4 1 Installation of Sensor Cartridge 5 4 2 Registration of the Sensor Cartridge 5 5 Calibration 6 5 1 Calibration Work 6 5 2 Exch...

Страница 3: ...C2 MSB2 and WSB2 devices 1 3 Special Mode See description of the SB2 MSC2 MSB2 and WSB2 devices 1 4 Sensor Element for Toxic Gases and Oxygen The sensor element is a sealed electro chemical cell with...

Страница 4: ...1186 1 27 Floor 0 2 0 3 m Hydrogen cyanide E1183 0 93 Ceiling 0 3 m Ethylene oxide E1199 1 56 Floor 0 2 0 3 m Ethylene E1189 0 97 1 5 1 8 m Formaldehyde E1185 1 04 1 5 1 8 m Carbon monoxide E1110 0 97...

Страница 5: ...stalled ex works Check Sensor Cartridge connector for correct engagement Calibrate if not already factory calibrated Within the first weeks after commissioning there may be a deviation in the sensor b...

Страница 6: ...libra tion interval in months1 Ammonia NH3 24 300 150 12 Chlorine Cl2 NO2 1 1 CL2 24 30 500 6 Hydrogen chloride HCL H2S 20 31 HCl 24 30 500 6 Hydrogen cyanide HCN H2S 15 25 HCN 24 30 500 6 Ethylene ox...

Страница 7: ...calibration has been successfully completed If you exit the menu without pressing Save the original gain data for the measured value calculation will continue to be used By limiting the gain factor ca...

Страница 8: ...epend on the sensor element and can be read from the table Calibration If there are different intervals always observe the shortest one Inspections and services must be documented The date for the nex...

Страница 9: ...ension at X4 16 28 V DC Pin 1 and 2 Polarity not correct at X4 Connect correctly Connector X4 not plugged in Check the plug Wire breakage Check the wiring Green LED doesn t flash PCB no address Check...

Страница 10: ...rent signal 50 k for voltage signal 4 20 mA or 2 10 V measuring range 3 2 4 mA or 1 6 2 V underrange 20 21 2 mA or 10 10 6 V overrange 2 mA or 1 V fault 21 8 mA or 10 9 V fault High SENSOR ELEMENT Gas...

Страница 11: ...1 F to 68 F for NH3 CO NO2 O2 Storage time 6 months Poisoning Electrochemical sensors are susceptible to poisoning by organic solvents and silicone vapours PHYSICAL Housing in plastic Polycarbonate Co...

Страница 12: ...1185 BX 0 10 n d 0 01 5 60 0 2 1 2 30 50 15 90 36 1 04 6 CO E1110 CX 0 150 2 0 1 5 10 4 0 4 0 4 20 50 10 95 72 0 97 12 CO E1110 EX 0 250 2 0 1 5 10 4 0 4 0 4 20 50 10 95 72 0 97 12 CO E1110 FX 0 300 2...

Страница 13: ...5000 0 10 12 10 30 20 0 20 0 1000 150 NH3 E1125 CX 10 0 100 0 100 0 200 0 5000 0 10 12 10 30 20 0 20 0 1000 150 NH3 E1125 DX 10 0 100 0 100 0 200 0 5000 0 10 12 10 30 20 0 20 0 1000 150 NH3 E1125 EX 1...

Страница 14: ...1 9004 0 Fax 49 8531 9004 54 Specification subject to change without notice MSR Electronic GmbH Buergermeister Schoenbauer Str 13 D 94060 Pocking www msr electronic de 10 Figures v Fig 1 Fig 3 PCB wit...

Страница 15: ...d 2 devices are installed in compliance with all national and local codes and OSHA requirements Installation should be implemented only by technicians familiar with proper installation techniques and...

Отзывы: