Zero calibration 

 

Apply  synthetic  air  (flow  rate  according  to  the  table  "calibration",  1  bar  ±  10%)  to  the  Sensor 
Cartridge.  

 

The current zero offset and the offset value of the first calibration is read by displaying.  

 

When the value is stable, confirm and the new zero offset factor is calculated.  

The  new  offset  factor  is  checked  for  plausibility  and  stored  in  the  buffer  memory.  The  current 
measured value is output with the new offset factor and the offset display is updated.  

 

With "Save" the new offset factor is written in the SC  memory, only  then the Zero calibration has 
been successfully completed. If you exit the menu without pressing "Save", the original offset data 
for the measured value calculation will continue to be used.  

With a zero reading > 10 % of measuring range during the zero calibration, zero calibration is not possible.  

Gain calibration  

 

Enter test gas concentration (value between 30 and 90 % of the measuring range)  

 

The current sensor element sensitivity is read by displaying.  

 

Apply test gas (flow rate according to the table "Calibration", 1 bar ± 10%) to the Sensor Cartridge.  

 

When the value is stable, confirm and the new gain factor is calculated. 

The  new  gain  factor  is  checked  for  plausibility  and  stored  in  the  buffer  memory.  The  current 
measured value is output with the new gain factor and the sensor element sensibility is updated.  

 

With "Save" the new gain factor is written in the SC memory, only then the gain calibration has been 
successfully completed. If you exit the menu without pressing "Save", the original gain data for the 
measured value calculation will continue to be used.  

By  limiting  the  gain  factor,  calibration  will  not  be  possible  any  more  when  the  sensitivity  of  the  sensor 
reaches a residual sensitivity of 40 %. Then the Sensor Cartridge has to be replaced  

For more information, see the user manual of the corresponding tool. 

5.2

 

Exchange of Sensor Cartridge 

 

Instead of the on-site calibration, the used SC can be replaced simply and conveniently by a calibrated one. 

The communication of the local bus (Sensor Cartridge <> PCB) is continuously monitored during 
operation and results in an immediate error message on the gas controller in case of fault  or 
interruption. When replacing the sensor unit, the communication of the local bus is interrupted 
when  unplugging  the  SC  connector  which  leads  to  an  immediate  triggering  of  the  error 
message. 

 

Disconnect the SC connector from the PCB or the remote PCB (error message will be activated). 

 

Loosen the locknut. 

 

Remove used SC. 

 

Take  calibrated  SC  out  of  the  original  packaging,  check  for  gas  type,  measuring  range  and  valid 
calibration date. 

 

Insert the SC and retighten with lock nut.  

 

Insert the SC plug into the socket at the PCB or remote PCB. Check plug for proper engagement. 

The local bus communication is automatically established and tested. At the same time the gas type and the 
measuring  range  of  the  "new"  SC  are  compared  with  the  data  stored  in  the  PCB.  If  they  match  and  the 
communication is correct, the error message will be automatically acknowledged in the Gas Controller.  

The  yellow  LED  of  the  PCB  flashes  with  a  pulse  of  1  sec.  as  long  as  the  SC  connector  is  disconnected 
(communication error). After the local bus communication has been re-established and the conformity test 
has been successful, the LED goes into flashing mode with 3 sec. pulse duration 

until the sensor’s warm

-up 

time is over.  

Содержание PolyGard 2

Страница 1: ...t to change without notice MSR Electronic GmbH Buergermeister Schoenbauer Str 13 D 94060 Pocking www msr electronic de PolyGard 2 Gard 2 Sensors for Toxic Gases Sensor Cartridge for Toxic Gases and Ox...

Страница 2: ...n SB2 MSB2 MSC2 WSB2 5 3 2 Terminal Connection MC2 5 4 Commissioning 5 4 1 Installation of Sensor Cartridge 5 4 2 Registration of the Sensor Cartridge 5 5 Calibration 6 5 1 Calibration Work 6 5 2 Exch...

Страница 3: ...C2 MSB2 and WSB2 devices 1 3 Special Mode See description of the SB2 MSC2 MSB2 and WSB2 devices 1 4 Sensor Element for Toxic Gases and Oxygen The sensor element is a sealed electro chemical cell with...

Страница 4: ...1186 1 27 Floor 0 2 0 3 m Hydrogen cyanide E1183 0 93 Ceiling 0 3 m Ethylene oxide E1199 1 56 Floor 0 2 0 3 m Ethylene E1189 0 97 1 5 1 8 m Formaldehyde E1185 1 04 1 5 1 8 m Carbon monoxide E1110 0 97...

Страница 5: ...stalled ex works Check Sensor Cartridge connector for correct engagement Calibrate if not already factory calibrated Within the first weeks after commissioning there may be a deviation in the sensor b...

Страница 6: ...libra tion interval in months1 Ammonia NH3 24 300 150 12 Chlorine Cl2 NO2 1 1 CL2 24 30 500 6 Hydrogen chloride HCL H2S 20 31 HCl 24 30 500 6 Hydrogen cyanide HCN H2S 15 25 HCN 24 30 500 6 Ethylene ox...

Страница 7: ...calibration has been successfully completed If you exit the menu without pressing Save the original gain data for the measured value calculation will continue to be used By limiting the gain factor ca...

Страница 8: ...epend on the sensor element and can be read from the table Calibration If there are different intervals always observe the shortest one Inspections and services must be documented The date for the nex...

Страница 9: ...ension at X4 16 28 V DC Pin 1 and 2 Polarity not correct at X4 Connect correctly Connector X4 not plugged in Check the plug Wire breakage Check the wiring Green LED doesn t flash PCB no address Check...

Страница 10: ...rent signal 50 k for voltage signal 4 20 mA or 2 10 V measuring range 3 2 4 mA or 1 6 2 V underrange 20 21 2 mA or 10 10 6 V overrange 2 mA or 1 V fault 21 8 mA or 10 9 V fault High SENSOR ELEMENT Gas...

Страница 11: ...1 F to 68 F for NH3 CO NO2 O2 Storage time 6 months Poisoning Electrochemical sensors are susceptible to poisoning by organic solvents and silicone vapours PHYSICAL Housing in plastic Polycarbonate Co...

Страница 12: ...1185 BX 0 10 n d 0 01 5 60 0 2 1 2 30 50 15 90 36 1 04 6 CO E1110 CX 0 150 2 0 1 5 10 4 0 4 0 4 20 50 10 95 72 0 97 12 CO E1110 EX 0 250 2 0 1 5 10 4 0 4 0 4 20 50 10 95 72 0 97 12 CO E1110 FX 0 300 2...

Страница 13: ...5000 0 10 12 10 30 20 0 20 0 1000 150 NH3 E1125 CX 10 0 100 0 100 0 200 0 5000 0 10 12 10 30 20 0 20 0 1000 150 NH3 E1125 DX 10 0 100 0 100 0 200 0 5000 0 10 12 10 30 20 0 20 0 1000 150 NH3 E1125 EX 1...

Страница 14: ...1 9004 0 Fax 49 8531 9004 54 Specification subject to change without notice MSR Electronic GmbH Buergermeister Schoenbauer Str 13 D 94060 Pocking www msr electronic de 10 Figures v Fig 1 Fig 3 PCB wit...

Страница 15: ...d 2 devices are installed in compliance with all national and local codes and OSHA requirements Installation should be implemented only by technicians familiar with proper installation techniques and...

Отзывы: