Zero calibration
•
Apply synthetic air (flow rate according to the table "calibration", 1 bar ± 10%) to the Sensor
Cartridge.
•
The current zero offset and the offset value of the first calibration is read by displaying.
•
When the value is stable, confirm and the new zero offset factor is calculated.
The new offset factor is checked for plausibility and stored in the buffer memory. The current
measured value is output with the new offset factor and the offset display is updated.
•
With "Save" the new offset factor is written in the SC memory, only then the Zero calibration has
been successfully completed. If you exit the menu without pressing "Save", the original offset data
for the measured value calculation will continue to be used.
With a zero reading > 10 % of measuring range during the zero calibration, zero calibration is not possible.
Gain calibration
•
Enter test gas concentration (value between 30 and 90 % of the measuring range)
•
The current sensor element sensitivity is read by displaying.
•
Apply test gas (flow rate according to the table "Calibration", 1 bar ± 10%) to the Sensor Cartridge.
•
When the value is stable, confirm and the new gain factor is calculated.
The new gain factor is checked for plausibility and stored in the buffer memory. The current
measured value is output with the new gain factor and the sensor element sensibility is updated.
•
With "Save" the new gain factor is written in the SC memory, only then the gain calibration has been
successfully completed. If you exit the menu without pressing "Save", the original gain data for the
measured value calculation will continue to be used.
By limiting the gain factor, calibration will not be possible any more when the sensitivity of the sensor
reaches a residual sensitivity of 40 %. Then the Sensor Cartridge has to be replaced
For more information, see the user manual of the corresponding tool.
5.2
Exchange of Sensor Cartridge
Instead of the on-site calibration, the used SC can be replaced simply and conveniently by a calibrated one.
The communication of the local bus (Sensor Cartridge <> PCB) is continuously monitored during
operation and results in an immediate error message on the gas controller in case of fault or
interruption. When replacing the sensor unit, the communication of the local bus is interrupted
when unplugging the SC connector which leads to an immediate triggering of the error
message.
•
Disconnect the SC connector from the PCB or the remote PCB (error message will be activated).
•
Loosen the locknut.
•
Remove used SC.
•
Take calibrated SC out of the original packaging, check for gas type, measuring range and valid
calibration date.
•
Insert the SC and retighten with lock nut.
•
Insert the SC plug into the socket at the PCB or remote PCB. Check plug for proper engagement.
The local bus communication is automatically established and tested. At the same time the gas type and the
measuring range of the "new" SC are compared with the data stored in the PCB. If they match and the
communication is correct, the error message will be automatically acknowledged in the Gas Controller.
The yellow LED of the PCB flashes with a pulse of 1 sec. as long as the SC connector is disconnected
(communication error). After the local bus communication has been re-established and the conformity test
has been successful, the LED goes into flashing mode with 3 sec. pulse duration
until the sensor’s warm
-up
time is over.