3.11
Time of Day Testing
This feature permits the device to be automatically calibrated on a user-defined interval. The most
common use of this feature allows the user to configure the ALTAIR 4XRand the GALAXY GX2
System to automatically calibrate a device prior to the start of work-shift. See the GALAXY GX2 Oper-
ating Manual ("Automated Testing Features" section) for a complete description of how to configure
the GALAXY GX2 for this mode.
On ALTAIR 4XR devices, the following settings must be configured using either MSA Link software or
the GALAXY GX2
Instrument Setup page:
•
For automated calibration testing, Calibration Due must be enabled and a non-zero Calibration
Interval must be entered for automatic calibration testing.
•
For automated bump testing, Bump Due must be enabled and a non-zero Bump Interval must be
entered for automatic bump testing.
Carefully follow all GALAXY GX2 set-up directions as described in the GALAXY GX2 Operating
Manual for proper setup.
Содержание ALTAIR 4XR
Страница 1: ...MSAsafety com Operating Manual ALTAIR 4XR Multigas Detector Order No 10175896 00 ...
Страница 52: ...ALTAIR 4XR 52 Appendix US 7 Appendix 7 1 Start Up Sequence Power on ...
Страница 53: ...ALTAIR 4XR 53 Appendix US 7 2 Fresh Air Setup FAS ...
Страница 55: ...ALTAIR 4XR 55 Appendix US ...
Страница 56: ...ALTAIR 4XR 56 Appendix US 7 4 Bump Test ...
Страница 57: ...ALTAIR 4XR 57 Appendix US 7 5 Options Setup ...
Страница 58: ...ALTAIR 4XR 58 Appendix US 7 6 Sensor Setup ...
Страница 59: ...ALTAIR 4XR 59 Appendix US 7 7 Calibrations ...
Страница 60: ...ALTAIR 4XR 60 Appendix US ...
Страница 61: ...ALTAIR 4XR 61 Appendix US 7 8 Alarm Setup ...
Страница 62: ...ALTAIR 4XR 62 Appendix US 7 9 Time and Date Setup ...
Страница 63: ...sales norrscope com ALTAIR 4XR 63 Appendix US 7 10 Bluetooth Setup ...