Operation
US
Attach the Calibration Cap
Span Calibration
Attach the calibration cap to the device:
(1)
Insert tab on calibration cap into slot on device.
(2)
Press calibration cap as shown until it seats onto
device.
(3)
Press both side tabs down onto device until they
snap into place.
(4)
Ensure that the calibration cap is properly seated.
(5)
Connect one end of the tubing to the calibration
cap.
(6)
Connect other end of tubing to the cylinder regu-
lator (supplied in the calibration kit).
To skip the Span procedure, push the
button.
If no button is pushed for 30 seconds, the device returns to Normal Operation mode.
(1)
Once the Zero is set, the
SPAN CAL?
screen displays.
(2)
Connect the regulator to the calibration
check gas cylinder.
(3)
Connect the appropriate calibration gas
to the device.
(4)
Attach the calibration cap (see chapter
3.10 "Attach the Calibration Cap")
(5)
Open the pressure reducer valve on test
gas cylinder.
(6)
Press the
q
button to calibrate (span)
the device.
SPAN
calibration starts.
(7)
Close the valve after
SPAN
calibration.
Содержание ALTAIR 4XR
Страница 1: ...MSAsafety com Operating Manual ALTAIR 4XR Multigas Detector Order No 10175896 00 ...
Страница 52: ...ALTAIR 4XR 52 Appendix US 7 Appendix 7 1 Start Up Sequence Power on ...
Страница 53: ...ALTAIR 4XR 53 Appendix US 7 2 Fresh Air Setup FAS ...
Страница 55: ...ALTAIR 4XR 55 Appendix US ...
Страница 56: ...ALTAIR 4XR 56 Appendix US 7 4 Bump Test ...
Страница 57: ...ALTAIR 4XR 57 Appendix US 7 5 Options Setup ...
Страница 58: ...ALTAIR 4XR 58 Appendix US 7 6 Sensor Setup ...
Страница 59: ...ALTAIR 4XR 59 Appendix US 7 7 Calibrations ...
Страница 60: ...ALTAIR 4XR 60 Appendix US ...
Страница 61: ...ALTAIR 4XR 61 Appendix US 7 8 Alarm Setup ...
Страница 62: ...ALTAIR 4XR 62 Appendix US 7 9 Time and Date Setup ...
Страница 63: ...sales norrscope com ALTAIR 4XR 63 Appendix US 7 10 Bluetooth Setup ...