ALTAIR
®
4XR
17
Description
US
Time Weighted Average (TWA page)
WARNING!
If the TWA alarm activates, leave the contaminated area immediately; the ambient gas concentration
has reached the preset TWA alarm level. Failure to follow this warning will cause over-exposure to toxic
gases and persons relying on this product for their safety could sustain serious personal injury or death.
The TWA icon (
chapter 2.3 "On-Screen Indicators") appears on the display to show the average
exposure since the device was turned on or since the TWA reading was reset. When the amount of
gas detected is greater than the eight-hour TWA limit:
•
Alarm sounds
•
Alarm LEDs flash
•
TWA icon flashes.
To reset the TWA:
(1)
Access the TWA page.
(2)
Press the
button.
The TWA alarm is calculated over an eight-hour exposure.
TWA calculation examples:
1 hour exposure of 50 ppm:
4 hour exposure of 50 ppm and 4 hour exposure of 100 ppm:
12 hour exposure of 100 ppm:
This page can be deactivated through the MSA GALAXY GX2 Test Stand or MSA Link software.
Time Display
Current time appears on the display in a 12-hour format by default.
A 24-hour format can be selected using MSA GALAXY GX2 Test Stand or MSA Link.
Date Display
Current date appears on the display in the format:
MMM-DD-YYYY
.
Discoverable Mode page
Allows the user to put the device into Bluetooth discoverable mode in order to pair with another device.
This page can be deactivated through the BT SETUP page.
(1 hour x 50 ppm) + (7 hours x 0 ppm)
= 6.25 ppm
8 hours
(4 hours x 50 ppm) + (4 hours x 100 ppm)
= 75 ppm
8 hours
(12 hours x 100 ppm)
= 150 ppm
8 hours
Содержание ALTAIR 4XR
Страница 1: ...MSAsafety com Operating Manual ALTAIR 4XR Multigas Detector Order No 10175896 00 ...
Страница 52: ...ALTAIR 4XR 52 Appendix US 7 Appendix 7 1 Start Up Sequence Power on ...
Страница 53: ...ALTAIR 4XR 53 Appendix US 7 2 Fresh Air Setup FAS ...
Страница 55: ...ALTAIR 4XR 55 Appendix US ...
Страница 56: ...ALTAIR 4XR 56 Appendix US 7 4 Bump Test ...
Страница 57: ...ALTAIR 4XR 57 Appendix US 7 5 Options Setup ...
Страница 58: ...ALTAIR 4XR 58 Appendix US 7 6 Sensor Setup ...
Страница 59: ...ALTAIR 4XR 59 Appendix US 7 7 Calibrations ...
Страница 60: ...ALTAIR 4XR 60 Appendix US ...
Страница 61: ...ALTAIR 4XR 61 Appendix US 7 8 Alarm Setup ...
Страница 62: ...ALTAIR 4XR 62 Appendix US 7 9 Time and Date Setup ...
Страница 63: ...sales norrscope com ALTAIR 4XR 63 Appendix US 7 10 Bluetooth Setup ...