Operation
US
Fresh Air Setup (FAS)
The FAS is for automatic Zero calibration of the device.
The Fresh Air Setup has limits. The zero of any sensor that is outside of these limits will not be adjusted
by the FAS command.
WARNING!
Do not activate the Fresh Air Setup unless you are certain you are in fresh, uncontaminated air; other-
wise, inaccurate readings can occur which can falsely indicate that a hazardous atmosphere is safe. If
you have any doubts as to the quality of the surrounding air, do not use the Fresh Air Setup feature.
Do not use the Fresh Air Setup as a substitute for daily calibration checks. The calibration check is
required to verify span accuracy. Failure to follow this warning can result in serious personal injury or
death.
Fig. 3
Fresh Air Setup
Fresh Air Setup (FAS)
The FAS screen is prompted.
If a battery charging cycle is interrupted before it is completed (4 hours for a fully
discharged battery), allow the device’s internal temperature to stabilize for 30 minutes
before performing a Fresh Air Setup.
Содержание ALTAIR 4XR
Страница 1: ...MSAsafety com Operating Manual ALTAIR 4XR Multigas Detector Order No 10175896 00 ...
Страница 52: ...ALTAIR 4XR 52 Appendix US 7 Appendix 7 1 Start Up Sequence Power on ...
Страница 53: ...ALTAIR 4XR 53 Appendix US 7 2 Fresh Air Setup FAS ...
Страница 55: ...ALTAIR 4XR 55 Appendix US ...
Страница 56: ...ALTAIR 4XR 56 Appendix US 7 4 Bump Test ...
Страница 57: ...ALTAIR 4XR 57 Appendix US 7 5 Options Setup ...
Страница 58: ...ALTAIR 4XR 58 Appendix US 7 6 Sensor Setup ...
Страница 59: ...ALTAIR 4XR 59 Appendix US 7 7 Calibrations ...
Страница 60: ...ALTAIR 4XR 60 Appendix US ...
Страница 61: ...ALTAIR 4XR 61 Appendix US 7 8 Alarm Setup ...
Страница 62: ...ALTAIR 4XR 62 Appendix US 7 9 Time and Date Setup ...
Страница 63: ...sales norrscope com ALTAIR 4XR 63 Appendix US 7 10 Bluetooth Setup ...