9010/20 LCD
Sensor data sheet DF-9500
with safety barrier
Part No.: D0742xxx
xxx – according to materials list
Terminal Block
9010/9020
External power
supply
Sensor Type DF-9500
Safety barrier
Caution:
Before connecting the measuring head, check hardware and software
settings.
Sensor type:
EC cell/active/2-lead/4–20 mA linear/current sink [2 wire active]
Connection data:
Supply voltage:
14 – 28 V DC
Maximum current draw:
50 mA
Power requirement:
1 W [excluding cable length]
Connection cable:
2-lead, 80 % shielded
Maximum resistance:
500 Ohm
Maximum cable length:
1.500 m [with 1.5 mm
2
lead cross-section]
Cable diameter:
8 – 12 mm
Permitted lead cross-section:
0.75 – 2.5 mm
2
Cable entry:
M20 x 1.5
Conditions for
use:
Assembly:
Wall-mounted
Protection:
IP 42 [sensor] IP 63 [housing]
Explosion protection:
Ex II 2G EEx ia IIC T6
Approvals:
DMT 01 ATEX E 152 X
Temperature:
min –20 °C to +45 °C [material-dependent]
Humidity:
5 – 95 % relative humidity; non-condensing
Pressure:
950 – 1,100 hPa
Weight:
approx. 0.65 kg
Dimensions:
100 mm x 95 mm x 165 mm
Содержание 9010 LCD
Страница 1: ...Operating Manual CONTROL UNIT MODELS 9010 and 9020 LCD Order No 10093590 02 ...
Страница 2: ...MSA AUER GmbH Thiemannstrasse 1 D 12059 Berlin Germany MSA AUER GmbH All rights reserved ...
Страница 84: ...Control Unit Models 9010 LCD and 9020 LCD 84 Appendix MSA GB ...
Страница 86: ...N o part no DATA DATE DESCRIZIONE DESCRIPTION PREPARATO PREPARED BY APPROVATO APPROVED BY REVISIONE REVISION ...
Страница 87: ......
Страница 88: ......
Страница 89: ......
Страница 90: ...No part no DATA DATE REVISIONI REVISIONS PREPARATO PREPARED BY APPROVATO APPROVED BY REVISIONE REVISION ...
Страница 94: ......
Страница 98: ......
Страница 99: ......