MSA AUER
MSA
Setting Parameters
Control Unit Models 9010 LCD and 9020 LCD
25
GB
The state of analogue output signals are as follows, depending on the preselected Access Code:
Access Code 1:
channel being calibrated as set in step P2 of Access Code 2
of the other channel: fluctuating
Access Code 2, 7, 123 and 223:
fluctuating for both the channels if the selection was made via Access Code 123, while they are
equal to 0 mA if the selection occurred via the ENTER button when the Control Unit is turned on.
Access codes 3 and 5:
fluctuating for both channels.
Access codes 4, 52 and 53:
-
selected channel: 0 mA
-
the other channel: fluctuating if the selection was made via Access Code 123, while it is equal
to 0 mA if the selection occurred via the ENTER button when the Control Unit is turned on.