Maintenance and Cleaning
SSR 90 [K60]
12
GB
4
Maintenance and Cleaning
4.1
Maintenance Instructions
To maintain readiness for use, the device must be inspected regularly, checked for
integrity, tightness and, if necessary, cleaned [do not use metal brushes!].
As the device contains a chemical which is sensitive to moisture, it is packed in an
airtight container and is sealed at the factory.
The frequency of inspection depends on the device load [
Æ
chapter 4.4].
For used and/or opened devices, the chemical canister must always be replaced,
reuse is not permitted. Internal checks and any repairs to the devies are to be car-
ried out by the MSA Customer Service or authorized service centers.
4.2
Return of Used Device
For return of used or opened devices, observe the following:
-
Pack the device and components in the device containers only after they have
cooled down and are dry.
-
No organic substances must enter the chemical canisters [e.g. solvents, oils,
etc.].
-
Observe transport regulations [
Æ
chapter 5].
If the complete device cannot be packed in the container, the following procedure
is required:
-
Remove the chemical canister from the device:
(1) Therefor loosen the clamp and remove the breathing bag and breathing tube.
(2) Pack the chemical canister inside the container and seal with the original lock-
ing assembly of the container.
Z
Do not use finely cutted filling material.
-
Place the other components of the device with the closed container in a carefully
packed package.
-
The components must be complete and undamaged.