Water Purification System                                                                                                      Instruction Manual 

  8 / 11 

password(In this process

, if password is not right, system will warn.) At last press “OK” key-press to switch 

to menu interface(picture 15), After switching to menu interface, through pressing “MENU” key-press, shift 
cursor to corresponding item menu, then press “OK” key-press to enter corresponding item. 

 

 

 

 

         

 

                         

picture 14                                                  picture 15 

(1).  Conductivity(TDS) parameter setting(picture 16): 

Under the menu interface(picture 15), press “MENU” key-press to shift the cursor to choose “TDS” 

item, then p

ress “OK” key-press to enter ”TDS” parameter setting(picture 16).   

Press “MENU” key-press to choose modified item, and press “      ” key-press to choose modified 
place, and press “      ” key-press to amend the parameter. When amending is ok, press “OK” 

key-p

ress for system’s confirmation. If back to main interface is necessary, press ”MENU” key-press 

to shift cursor to icon at the right bottom of the LCD and press “OK” key-press to return to main 

interface. 

In the following item parameter setting process, the method is equivalent to the 

conductivity(TDS) parameter setting’s.

 

TDS1: inlet water conductivity test. 

Max: inlet water conductivity’s upper limit alarm parameter of TDS1. 

Min: inlet water conductivit

y’s lower limit alarm parameter of TDS1. 

Remark:

   

TDS1(hysteresis of inlet water conductivity =Max-Min) upper and lower alarm test only work when 
power is on. When current test parameter > TDS1’s Max, the current test parameter will display and 

system will 

warn. System will stop warning when current test parameter <TDS1’s Max.   

TDS2: RO water conductivity test. 

Max: RO water conductivity’s upper limit alarm parameter of TDS2. 
Min: RO water conductivity’s lower limit alarm parameter of TDS2. 

Remark: When cur

rent test parameter of RO water > TDS2’s Max, the current test parameter will 

display and system will warn. System will stop warning when current test parameter <TDS2’s Max.   

EC: electrode parameter setting (Parameter setting is OK in factory. Additional parameter setting is 

unnecessary.) 

TCQ: temperature compensation parameter setting(Parameter setting is OK in factory. Additional 

parameter setting is unnecessary.) 

 

 

 

                                  picture 16                                                picture 17 

(2).  Resistivity parameter setting 

Under the menu interface(picture 15), press “MENU” key-press to shift the cursor to choose “Resist” 
item, then press “OK” key-press to enter ” Resist”” parameter setting(picture 11). 
Max: resistivity’s(Ultrapure polishing resin cartridge’ life

upper limit alarm parameter. 

Min: resistivity’s(Ultrapure polishing resin cartridge’ life

lower limit alarm parameter 

Remark:   

The “Min” determines alarm of Ultrapure polishing resin cartridge’ life. When current test parameter of 

Содержание Master-S UVF

Страница 1: ...ater Purification System Instruction Manual Model Master S UVF The instruction manual should be read before installation Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC...

Страница 2: ...day Model Master S UVF Quality of Ultrapure water Resistivity 18 2 M cm TOC 3ppb Endotoxin 0 001EU ml Bacteria 1 CFU ml Particle 0 1 m 1 ml Heavy metal ion 0 1ppb Quality of RO water TDS ppm mg l TDS...

Страница 3: ...uality easy put adapter Proper usage is as follows Connect the tube press the oval cap of the interface strongly then insert the tube to the bottom of adapter Take off the tube press the oval cap of t...

Страница 4: ...of machine 3 Connect to RO drain Use 1 4 PE tube with a suitable length One side connects with the interface with black label marked To drain at the back of machine and the other side is directed to g...

Страница 5: ...uatic products feeding inlet water for ultra pure water machine inlet water for sterilizer t h chamber Terminal filter Tap water Tap water adapter 2 Tap water adapter 1 3 direction links Plastic ball...

Страница 6: ...Ultrapure water higher quality water than RO water Standby When RO water and Ultrapure water is not for use the system will be in standby state The system still produce RO water to store in the water...

Страница 7: ...xecute corresponding function UP key press open or close Ultrapure water valve Remarks The initial code is 1000 After pressing MENU key press please in code will appear Inputting the code 1000 through...

Страница 8: ...until the state of inlet water is ok And system will produce pure water again C When system is under working conditions it will warn for no inlet water or low pressure of inlet water The sound and pi...

Страница 9: ...r limit alarm parameter of TDS1 Remark TDS1 hysteresis of inlet water conductivity Max Min upper and lower alarm test only work when power is on When current test parameter TDS1 s Max the current test...

Страница 10: ...system s initial code is 1000 Under the menu interface picture 15 press MENU key press to shift the cursor to choose Code item then press OK key press to enter Code setting picture 19 Press key press...

Страница 11: ...trapure polishing resin cartridge 4 About 4000 liter pure water UF 5000D 5000 doulton uf cartridge 1 TF 0 45 0 1 m S 0 45 0 1 m terminal filter 1 Remarks Worse inlet source water quality or big dosage...

Страница 12: ...side water circuit breaker and replace it if it goes wrong B Pressure of inlet water is higher then the set pressure of high pressure switch Please regulate the pressure of inlet water no more then 2k...