  10 / 11 

Measure unit: TDS(total dissolved solid, ppm) 

 

The third, Resist, 

monitors Ultrapure 

water’s quality (Ultrapure water). 

Measure unit: Resistivity (

MΩ.cm) 

Remarks: 

 

Conversion relations between TDS and conductivity rate(

µs/cm

): 

If TDS<50ppm. 

conductivity rate(

µs/cm)

 ≈

TDS×2

 

If TDS>200ppm. 

conductivity rate(

µs/cm)

 ≈

TDS×(1.5~1.7)

 

 

If TDS2 of RO water < TDS1 of inlet tap water×10%(under normal conditions, TDS2 of RO water is around 

TDS1 of inlet tap water×5%), RO membrane is ok. If TDS2 of RO water > TDS1 of inlet tap water×10%, 

Replacing RO membrane is essential for proper pure water. 

 

9.

 

  Consumables 

Item No. 

Commodity 

Quantity/set 

Suggested replacement term 

PC-M-PP 

5μm spun fiber filter 

About 2-6 months 

PC-M-KDF 

Kinetic degradation fluxion filter 

About 4-6 months 

PC-M-AC-G 

Granular active carbon filter 

About 4-6 months 

RO-100GPD 

100 gpd ro membrane 

About 12-24months 

LAMP-(185nm&254nm)-10W-M 

Double  wavelength(185&254)nm  uv 

lamp 

About 9000 hours 

PTC-AC-HZB 

Post active carbon filter 

About 9000 liter pure water 

PTC-UPPR-M 

Ultrapure polishing resin cartridge 

About 4000 liter pure water 

UF-5000D 

5000 doulton uf cartridge 

TF-(0.45+0.1

)μm-S 

(0.45+0.1

)μm terminal filter 

Remarks: 

 

Worse inlet source water quality or big dosage will reduce the life of cartridge above. 

 

10.

 

Trouble Diagnosis 

. Motor failure 

A. Check whether the pressure of the tap water is high enough. When pressure is low, low pressure switch 

will cut off power. 

B. Check whether the high pressure switch or water level controllor is normal. 

C. Check whether special PP filter or Active carbon is blocked which cause the overload of the motor. 

D. Check whether control panel is OK (for full automatic model) 

 

. Pump Leaking 

A.  Check  whether  special  PP  filter  or  Active  carbon  is  blocked  which  cause  the  overload  of  the  motor. 

Then high pressure lead to leaking. Resolution: change special PP filter or Active carbon and repair the 

pump. 

B. The expiring of the pump rubber membrane or the dirt will cause leaking too. Please repair the pump 

and pay attention to checking the LV. 1-special PP cotton. 

C.  The  block  of  the  waste  water  proportion  vessel  will  cause  the  block  of  the  reverse membrane.  Then 

pressure  increase.  Water  with  high  pressure  will  break  the  pump  rubber  membrane,    then  causes 

leaking. Change relative components will be ok. 

 

Содержание Master-S UVF

Страница 1: ...ater Purification System Instruction Manual Model Master S UVF The instruction manual should be read before installation Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC...

Страница 2: ...day Model Master S UVF Quality of Ultrapure water Resistivity 18 2 M cm TOC 3ppb Endotoxin 0 001EU ml Bacteria 1 CFU ml Particle 0 1 m 1 ml Heavy metal ion 0 1ppb Quality of RO water TDS ppm mg l TDS...

Страница 3: ...uality easy put adapter Proper usage is as follows Connect the tube press the oval cap of the interface strongly then insert the tube to the bottom of adapter Take off the tube press the oval cap of t...

Страница 4: ...of machine 3 Connect to RO drain Use 1 4 PE tube with a suitable length One side connects with the interface with black label marked To drain at the back of machine and the other side is directed to g...

Страница 5: ...uatic products feeding inlet water for ultra pure water machine inlet water for sterilizer t h chamber Terminal filter Tap water Tap water adapter 2 Tap water adapter 1 3 direction links Plastic ball...

Страница 6: ...Ultrapure water higher quality water than RO water Standby When RO water and Ultrapure water is not for use the system will be in standby state The system still produce RO water to store in the water...

Страница 7: ...xecute corresponding function UP key press open or close Ultrapure water valve Remarks The initial code is 1000 After pressing MENU key press please in code will appear Inputting the code 1000 through...

Страница 8: ...until the state of inlet water is ok And system will produce pure water again C When system is under working conditions it will warn for no inlet water or low pressure of inlet water The sound and pi...

Страница 9: ...r limit alarm parameter of TDS1 Remark TDS1 hysteresis of inlet water conductivity Max Min upper and lower alarm test only work when power is on When current test parameter TDS1 s Max the current test...

Страница 10: ...system s initial code is 1000 Under the menu interface picture 15 press MENU key press to shift the cursor to choose Code item then press OK key press to enter Code setting picture 19 Press key press...

Страница 11: ...trapure polishing resin cartridge 4 About 4000 liter pure water UF 5000D 5000 doulton uf cartridge 1 TF 0 45 0 1 m S 0 45 0 1 m terminal filter 1 Remarks Worse inlet source water quality or big dosage...

Страница 12: ...side water circuit breaker and replace it if it goes wrong B Pressure of inlet water is higher then the set pressure of high pressure switch Please regulate the pressure of inlet water no more then 2k...

Отзывы: