Water Purification System Instruction Manual
9 / 11
resistivity <
The “Min”, system will warn to replace the Ultrapure polishing resin cartridge. When
current test parameter of resistivity <
The “Min”, system will stop warning.
EC: electrode parameter setting (Parameter setting is OK in factory. Additional parameter setting is
unnecessary.)
(3).
Parts’ life setting / parts’ runtime clear
Under the menu interface(picture 15), press “MENU” key-press to shift the cursor to choose “Parts”
item, then press “OK” key-press to enter ”Parts” setting(picture 18).
Pre-treat: pre-
treat’s lift setting. (range: 0-9999 hours)
UV lamp: UV lamp’s life setting. (range: 0-9999 hours
picture 18
Pre-treat runtime: clr - Pre-treat runtime clear.
UV lamp runtime: clr - UV lamp runtime clear.
(press “MENU” key-press to shift the cursor to choose corresponding “clr” icon, then press “OK”
key-press to clear runtime. And system will restart timing)
(4). Code setting
All system’s initial code is “1000”
Under the menu interface(picture 15), press “MENU” key-press to shift the cursor to choose “Code”
item, then press “OK” key-press to enter ”Code” setting(picture 19). Press “ ” key-press to choose
modified place, and press “ ” key-press to amend the parameter. After amending code, system will
use new code for entering menu interface.
picture 19
(5). RO water, Ultrapure water timing dispense setting
Under the menu interface(picture 15), press “MENU” key-press to shift the cursor to choose
“dispense” item, then press “OK” key-press to enter ”dispense” setting(picture 20). 2 kind dispense
mode is optional-
“Manual” and “Auto”. When system enters “Auto” dispense mode, timing setting is
viable.
UP out: Ultrapure
water timing dispense setting (only for “Auto” dispense mode)
RO out: RO water timing dispense setting (only for “Auto” dispense mode)
picture 20
8.
Water Quality Test
The system has 3 monitor sensors of water quality measuring.
The first, TDS1,
monitors feed water’s quality (tap water, inlet water).
Measure unit: TDS(total dissolved solid, ppm)
The second, TDS2,
monitors reverse osmosis water’s quality(RO water).