WARNING
Be careful when you clean the infrared lens, the lens has a delicate anti-reflective coating. Do not clean the
lens too vigorously; this can damage the anti-reflective coating.
WARNING
The Product allows for field replacement of the lens. Exchanging the lens should be done with great caution
as after the removal of the lens the IR detector is exposed until the new lens is installed.
While exposed, avoid any contact with the detector. Touching, cleaning or blowing the detector could
cause permanent damage. Replacement of lens should be done in a dry area, using clean hands with
maximum care.
WARNING
Do not use pressurized air from pneumatic air circuits in a workshop to remove dust from the detector. This
air contains oil mist intended to lubricate pneumatic tools and is too high of a pressure, this can damage the
detector and lens.
WARNING
The device is designed to use 5V DC power supply only, fed thru its micro USB connector. Using any other
type of power source is dangerous and may lead to damage.
WARNING
The Product is provided with appropriate USB-OTG cables. Use only the supplied USB OTG cables. Using any
other types of cables might not work at all, may fail, and even damage the equipment.