34
Analysis setup includes
Temp range
,
IR lens
,
Temp Unit
,
Ambient
Temp
,
Reference
,
Ref Temp
. press the
、
button to choose
different parameter, press the
、
button to adjust parameter value.
Temp Range
Select temp range bases on measuring target.
IR-CAM-4
00
/600
has a standard temp range of -40
℃
to 600
℃
, extra
temp range of -40
℃
to 1200
℃
.
IR Lens
For
IR-CAM-400/600
infrared camera, there are 3 options: 38mm
(standard lens), 18.5mm (wide-angle lens), and 65.2mm (telephoto lens);
Press
or
button to select correct IR lens for best measurement
result after installing optional IR lens,
The camera may need 20 seconds to readjust a new lens before it
can work properly. System not have fusion and PIP mode for telephoto
lens.
Temp Unit
Press
or
to select Temperature Unit from , , and K.
=1.8* +32
,
K=273.15+
Ambient Temp
For more accurate measure result, set Ambient Temp to correct
environmental temperature. Press
or
once to change Ambient
Temp by 0.1
℃
. Long press
or
will change Ambient Temp by 1
℃
interval.
Reference
Set reference spot or area to display measure results as relative value.
There are five options:
Off, Ref Temp, Spot1~Spot10, Rect1~Rect5.
Choose the “Off” option to show original temperature and choose other
options to show relative value.
Содержание IR-CAM-400
Страница 2: ......
Страница 4: ......
Страница 40: ...32 1 2 PIP 1 4 PIP 3 4 PIP ...
Страница 50: ...42 The effect drawing of indoor pipe heat leakage The effect drawing of indoor corner air leakage ...
Страница 55: ......
Страница 56: ...IR CAM 400 600 of Uncooled Focal Plane Infrared Cameras IRSee Report Software User Manual ...
Страница 57: ......
Страница 69: ...10 ...