27
temperature reading. During calibration, if temperature reading is higher
than actual target temperature, change ratio value lower, and vise versa.
Press
or
once to change ratio value by 0.01. Long press
or
will change ratio value by 0.1 intervals.
Adjust Temp
Under certain condition, it is necessary to adjust temperature reading by a
pre-define value. Normally this option should be left unchanged. Press
or
once to change by 0.1
O
C. Long press
or
will change this
value by 1
O
C interval.
Save Setting
All parameter changes must be saved before power off device. Otherwise,
changes will not be saved for next power on.
5.2.3 Image
There are 5 items: Image Setup, Analysis Setup, Image Reverse, Clear
Screen, and L&S (Auto/Manual) in Image Menu.
Image sub-menu