Brief, intense flashes at
rapid set intervals:

The RF stick is out of range or has been
switched off. If no burner is detected
after 40 flashes, the foot pedal
switches itself off.

Intermittent lighting up
of the LED, duration approx. 1 sec.:

batteries low 

(see paragraph 8)

Slight glow:

Programming in progress 

(see paragraph 2.4)

2.4 Foot pedal programming:

The laboratory burner and foot pedal can be newly programmed to each other if the signal
is interrupted or in cases of parallel use of several radio systems. For programming of the
radio channel, place the laboratory gas burner close to the wireless foot pedal (max. 30
centimetres). Programming has a restricted range to avoid inadvertent re-programming of
other burners. (In cases where undesired re-programming of a neighbouring burner occurs,
switch it off during the programming procedure.)

- Programming procedure:
Attention: Adhere to the sequence for the programming procedure. Close the gas
regulator (2) by turning it clockwise to avoid inadvertent activation of the flame.

1. Switch off laboratory gas burner.

2. Switch on foot pedal by operating briefly or press again briefly.

3. Switch the laboratory gas burner.

4. Shortly afterwards, press the programming key 

(20)

 on the foot pedal for approx. 1 second

to start the programming procedure. The function LED 

(20)

 will now start to glow. This procedure

can last for between 0.5 and 30 seconds. As soon as the laboratory burner and the foot
pedal have been successfully programmed, the glowing is switched off and the function
LED

 (20)

 switches to brief, faint flashes.

Note: The laboratory burner and the foot pedal have already been programmed to each
other by the manufacturers.

2.5  Operation: Application programs

- BUTTON StartStop:

 The flame is ignited by operation of the function knob 

(1)

. The flame

is extinguished after renewed actuation of the function knob 

(1)

.

- PEDAL Standard:

The flame is ignited by operation of the foot pedal. The foot pedal remains depressed for the
duration of use. The flame is extinguished upon release of the pedal.

- PEDAL Start-Stop:

The flame is ignited by operation of the foot pedal. The flame is extinguished after renewed
actuation of the foot pedal. Additionally the flame is automatically extinguished when the
burning timer has expired after 60 min. Alternatively the flame can be extinguished by a short
push on the function knob.

2.6 Operation: Switch-off

The unit can be turned off by pushing the function knob 

(1)

 for more than 2 seconds.

The foot pedal is deactivated automatically.

Содержание Fuego SCS basic

Страница 1: ......

Страница 2: ...vice for inoculation loop holder 12 Burner head screw 13 Monitor electrode 14 Ignition electrode 15 Active nozzle 16 Nozzle holder for alternative gas 17 Cover of the burner shaft 18 Retaining screw for cover of the burner shaft 18a Position scew for the cover 19 Guide slots for tilt adjustment 4 6 2a 2 5 7 1 7a 3 14 13 12 8 9 10 11 16 17 18 15 19 19 18a 20 Function LED and programming key 21 Batt...

Страница 3: ...hose clamps are required All gas connections must be adequately tightened with two wrenches Ensure gas proofness with a suitable test fluid equipment DO NOT seal the swivel nut with Teflon tape etc Keep hands or other parts of the body away from the burner orifice 7a Do not operate the unit near flammable liquids or hazardous materials Unattended operation of the unit is not permissible Always wor...

Страница 4: ... the same way and exchange the nozzles ATTENTION Before mounting a nozzle check the O Ring 20 Replace the sealing if damaged or worn Art No 8 000 010 Now you are ready to connect the gas supply to the gas inlet 10 The correct pressure for natural gas is within the range of 18 25 mbar for propane butane gas 28 57 mbar Only use DVGW or other gas approved safety tubings with thread or tubing connecto...

Страница 5: ...switched off The radio system can be used without changing any further settings as the laboratory gas burner and the wireless foot pedal have already been programmed to each other gas burner and foot pedal programming see paragraph 2 4 by the manufacturers Attention The laboratory gas burner is started as soon as the foot pedal is operated and the LEDs 3 and 4 stop blinking Attention Flame The cor...

Страница 6: ...gas burner 4 Shortly afterwards press the programming key 20 on the foot pedal for approx 1 second to start the programming procedure The function LED 20 will now start to glow This procedure can last for between 0 5 and 30 seconds As soon as the laboratory burner and the foot pedal have been successfully programmed the glowing is switched off and the function LED 20 switches to brief faint flashe...

Страница 7: ... 18 25 mbar Nozzle P propane butane gas 47 57 mbar Flame failure Green LED Button Standard or StartStop blinks 3x This signal indicates a malfunction if the flame is extinguished by external factors and fails to reignite within 5 sec In case of flame failure check the burner head 7 for possible clogging and verify the correct input pressure of the gas supply In case of this malfunction the gas sup...

Страница 8: ...ion is performed in the reverse sequence The dismounted burner head can be even dismantled into the individual components for in depth cleaning Unscrew both screws 12a and take off the base plate 12b of the burner head which was fixed by the two screws 12a After the base plate is removed both electrodes can be pulled out for seperate cleaning Reinstallation is performed in the reverse sequence 5 2...

Страница 9: ... batteries Change both mignon batteries AA in the battery holder in the wireless foot pedal 21 Pay attention to correct insertion in accordance with the batteries polarity see label in the battery holder of the foot pedal 9 Warranty All Fuegos are covered under our two year manufacturer warranty against any manufacture defects in material and workmanship The 05 warranty guarantees all Fuegos under...

Страница 10: ...Notes ...

Страница 11: ... cannot encircle the electrodes correctly Clean this areas with a brush The burner head can be cleaned with customary commercial disinfectants or it can be sterilized in an autoclave or washed in a dishwasher In operating mode Button or Pedal StartStop flame burns 30 seconds only BHC time limit is active LED Burner head hot BHC is blinking The burner head is clogged and must be cleaned see paragra...

Страница 12: ...his message indicates that the burner head is removed or not mounted correctly Reinstall burner head and reset the unit by a long push on the function knob 1 see paragraph 4 Amber LED Burner head HOT BHC is on permanently Residual heat display is active Attention DO NOT TOUCH the burner head see paragraph 3 LEDs Standard 3 and Start Stop 4 are blinking alternately Wireless foot pedal out of range ...

Страница 13: ...es Flame temperature 1350 C on liquid gas 1300 C on natural gas E Temperature threshold level 1 kW liquid gas 1 kW natural gas Electrical Power consumption 2 VA Power connection 100 240 V 50 60 Hz max 0 3 A 9 V DC 1 A Power supply foot pedal 2 x AA Mignon Mechanical Casing and operating controls stainless steel glass UV and solvent resistant Burner head removable and decomposable stainless steel C...

Страница 14: ...rd ElectricalEquipment classB table4 GenericImmunityStandard Industrialareas continuous unmonitored operation 1 2 EN61000 3 2 2000 Limitsforharmoniccurrentemissions classA 1 3 Protectionrequirementsconcerningelectromagneticcompatibility 3 1 2 Article3 1 b Electromagnetic compatibility and Radio spectrum Matters ERM ElectroMagnetic Compatibility EMC standard for radio equipment and services EN30148...

Отзывы: