7a
7
12
12a
12b
12b
offener
Brennerschacht
18a
17
18
18a
23
5. Cleaning and sterilizing:
Allow sufficient time for burner orifice (
7, 7a
) to cool down before disassembling or cleaning
the burner head. Check the unit is disconnected and that the gas supply is turned off at the
mains. The burner can be cleaned with customary commercial disinfectants. Additionally, it
is possible to remove the burner head and to clean it separately. The stainless steel and glass
construction allow 100% UV-radiation sterilization and short time surface flame sterilization.
Attention: Because of the connectors at the back of the unit the backside should not be
sterilized with a flame.
5.1 Burner head disassembly and cleaning:
Allow sufficient time for burner orifice (
7,7a
) to cool
down before disassembling or cleaning the burner
head. Check the unit is turned off, that the gas supply
is turned off at the mains. Clean the burner head with
customary commercial disinfectants, sterilize it in
an autoclave or wash it in a dishwasher. To remove
the burner head proceed as follows:
Unscrew the burner head screw (
12
) completely with
the included screwdriver. Turn approx. 8 revolutions to
the left.
Now remove the burner head from the device by pulling it upwards.
Reinstallation is performed in the reverse sequence.
The dismounted burner head can be even dismantled into the individual components for in-
depth cleaning: Unscrew both screws
(12a)
and take off the base plate
(12b)
of the burner
head which was fixed by the two screws
(12a)
. After the base plate is removed both electrodes
can be pulled out for seperate cleaning. Reinstallation is performed in the reverse sequence.
5.2 Burner shaft cleaning:
Unscrew the screw
(18)
completely with the
included screwdriver. Take off the cover
(17)
of the burner shaft. Now the burner shaft can
be cleaned or solid substances which have
fallen into the unit can be removed.
Reinstallation is performed in the reverse
sequence. Take care that the notch of the
cover fits to the screw
(18a).
6. Turbo flame:
If the cover of the burner shaft
(17)
is removed
the flame is extremely firm and consistent.
To take off the cover of the burner shaft unscrew
the screw
(18)
completely with the included
screwdriver. With an open burner shaft the intensity
of the flame cannot be adjusted by the air knob
(2a)
any longer. During the use of the turbo flame most
of the needed air is taken inside through the open
burner shaft. Remounting the cover of burner shaft.
(see paragraph 5.2)
7. Tilt adjustment:
Insert the tilt adjustment
(23)
into the slots
(19)
at the bottom of the unit. The tilt-adjustment can
be used to the left or right side to protect the
burning chamber from contamination when working
with liquids.