USER MANUAL
S E M S
- 35 -
5. APPENDIX A Communication _Protocol
Protocol is implemented on RS232 set as follows:
Baud Rate
115200
1 Start Bit
1 Stop Bit
No Parity
Semantic:
For BASIC REQUEST and PROPRIETARY REQUEST commands are made of
ASCII string delimited by the character
“*”
(0x2A) and the length is four characters
Replies are terminated with <CR><LF> (\r\n)
Applicable with the versions below or following:
-
ver. FW B.37
-
ver. FPGA 0x14
-
ver. MSP 430 2.07
Below, the
float
values represent the numbers in “Little Endian” notation:
-
every
float
will be contained in 4 bytes
-
every
unsigned int
will be contained in 2 bytes
BASIC REQUEST COMMANDS
Commands Description
IDN?
This command *
IDN?
gives version and date of the firmware, followed by
[0D][0A]
Example of reply: IDN=SEP-FW - B.37 20/10/17[0D][0A]
For a correct configuration of the units use in this sequence the command setting:
*IDN?
SENS:FREQ:STOP:<value of frequency>
SENS:FREQ:STAR:<value of frequency>
SENS:BAND:<filter RBW>
SENS:DETE:<detector type>
SENS:HOLD:<time for the hold>
SENS:AXIS:<name of axis>
DATA:ISTA:VERB
For more informations about commands parameter see below.
Содержание SEP
Страница 16: ...USER MANUAL S E M S 16 c Link d Report e End Figure 6 Installer MPB SEP ...
Страница 25: ...USER MANUAL S E M S 25 Figure 19 Manage Configurations ...
Страница 44: ...USER MANUAL S E M S 44 Select Add a device Bluetooth Figure 33 Select RNBT xxxx ...
Страница 45: ...USER MANUAL S E M S 45 Figure 34 Accept the bluetooth association by clicking on connect Figure 35 ...
Страница 47: ...USER MANUAL S E M S 47 Figure 37 ...