5.4 LPA3 removal and product maintenance
When removing the LPA3 from the system ensure the system pressure is shut off from the LPA3.
• Open the flush valve by selecting the manual flush button on the display
• Disconnect the high-pressure hydraulic hose from the M16x2 test point on the system
• Disconnect the hose from the particle analyser HP connector
• NOTE: there may be some residual oil in the hose when it has been disconnected, this must be cleaned and disposed of
safely
• Remove the waste hose from the waste connector by pulling back the outer collar.
• Allow any fluid to drain from the hose and then remove from the waste vessel
• NOTE: Ensure any spillages are cleaned up and that all fluids are disposed of in accordance with local legislations
• Wipe away any residual oil from around the connectors in the LPA3 bulkhead and then replace the caps
5.5 Disposal
All LPA3 products are sent in a cardboard box with appropriate protective packaging and these should be recycled accordingly
where possible.
Fluids used with the LPA3 should be fully drained and disposed of according to EU waste framework directive and ISO 44001
Environmental Management.