Section 5
5-3
Resting ECG:
(optional - requires 12-lead
Interpretation)
SpO
2
:
(optional - requires T12S)
Missing QRS, High Pause Rate, Irregular Rhythm, Low QRS voltage, Pacemaker Output
Failure, High Output Failure Rate, Pacemaker Non-Capture, High Non-Capture Rate
Low or information only priority/On or Off (user selectable)
QTc High: ........................... ..300-500 msec
QTc-Bazett High:................ ..300-500 msec
QTc-Fridericia High: .......... ..300-500 msec
High, medium or low priority/On or Off (user selectable)
SpO
2
Low ............................. Medium priority, 70-98% saturation
Desaturation: ......................... High priority, 10% below SpO
2
low limit
Pulse rate low ........................ Medium or Low priority, 20-100 pulse rate
Pulse rate high: ..................... Medium or Low priority, 50-250 pulse rate
Technical Alarms:
Information only
:
Battery Low, Electrode impedance test, 12-lead ECG taken, Synchronizing clocks, No AC-
Power, System battery power low
Low priority:
Printer error, Printer off-line, Storage space < 24hr, Deleting old sessions, Hard disk full,
Storage system unreachable, Repairing database
Low or information only (user selectable):
ECG Leads Off, ECG Noise, No Radio Signal, Radio Interference.
Medium, low, or information only/On or Off (user selectable):
Patient Call
Medium priority:
No Heart rate, ECG monitoring disabled, Alarm engine fault
Protocol events:
(optional)
Information only
:
Protocol completed, Protocol aborted
Medium, low or information only/On or Off (user selectable):
Step change, Phase change (user-selectable texts)
ALARMS OFF:
Information only with audible reminder (user selectable)
Storage
Period:
0 to 600 days (optional)
Continuous:
12 lead ECG at 500 Samples/s @ 2.5
V LSB (optional)
12 lead average waveform:
every 5 s: 500 Samples/s @ 2.5
V LSB (optional)
Trends:
Heart rate, Ventricular rate, R-on-T rate, Couplet rate, Pause rate, Pacemaker Non-capture
and Output Failure rates, PR-interval, QRS duration, QT interval QTc interval (
Fridericia), SpO
2
, 12-lead ST
Alarms:
Physiological alarms, clinical alarms, protocol events, 12-lead ECG printout events
RR-intervals and beat labels:
2 ms resolution
Configuration
Storage Configurations:
1) Storage on local workstation
2) Storage on redundant (RAID1+0) server
Component Configurations:
1) Receiving workstation with support for up to 16 channels
2)
Control workstation with support for up to 32 channels
3)
Control and receive on one workstation
4)
Control, receive, and storage on one workstation
5)
Storage on redundant (RAID1+0) network server
Maximum number of telemetry
channels per system:
Up to 128 channels per Surveyor Central system (depending on frequency band, 2500 MHz,
608 MHz or 915 MHz; more than one storage station is required above 64 patients),
NOTE:
Surveyor Central components are designed to share an independent physical
network. Connection to an external network is through an optional VPN security router.