14. Power Up and Power Down Procedure:
To Power down Surveyor Central Telemetry System proceed as follow:
1. First, power off the Control Nodes (SCV1 & SCV2) normally located in Nurses Station:
Stop monitoring of all patients.
Select the
Config
tab at the bottom right hand corner of the screen.
Password is
mortara or Facility’s password
.
Select
Shutdown Computer
.
2. Next, power off the Telemetry Nodes (SCT1 & SCT2) normally located in IT closet:
Expand the
Guardian
Process.
Right click on Guardian window and a pop up asking for the password will appear, password is
mortara
, and select
Shutdown System.
3. The last step in shutting down the Surveyor Telemetry Central is to shut down the Storage Node
(SCS1) also located in IT Closet:
Expand
Guardian
.
Right click on Guardian window and a pop up asking for the password will appear, password is
mortara
, and select
Shutdown System.
To Power On Surveyor Telemetry Central System proceed as follow:
1. Power on the system in the following order
Storage Node (SCS1)
, then the
Telemetry Nodes
(SCT1 & SCT2)
and finally the
Control Nodes (SCV1 & SCV2).
Note:
Powering off and on of the Surveyor Telemetry Central system has to be done in the exact orders as
stated above. Failure to follow instructions will cause unwanted problems. Disregard the procedure if your
Surveyor Telemetry Central System is a stand-a-lone system.
15. Repeater Node not working:
No Tracings seen on screen:
Check IP address:
1. Right click on Repeater shortcut Icon and select Properties
2. In the Shortcut Tab, verify the Target line lists the correct IP
3. Verify that all processes running on Guardian has a green dot next to them
Verify the channels are selected:
1. Click on CONFIG and enter password.
2. In the bed list make sure the beds you want are checked and view leads are selected.
16. Reviewer Node not working:
Verify all network cables are connected
Verify IP address is correct
Verify that all processes running on Guardian has a green dot next to them
Verify dongle has correct permissions
Check that the correct Data Source Type is selected
Verify location of Data Source Type is correct
If viewing Online Database, verify IP address is correct