WHITEHALL MANUFACTURING
•
P.O. BOX 3527
•
City of Industry, CA 91744-0527 U.S.A.
Phone (800) 782-7706
•
(626) 968-6681
•
Fax (626) 855-4862
•
www.whitehallmfg.com
23
Classification
Electric Shock Protection
Class I
Applied Parts
None
Ingress Protection
IP24
Mode of Operation
Continuous
Specifications
Mechanical
Water capacity
Liters (US Gallons)
P-10
P-15
P-22
S-85
S-90
S-110
SB-100
54.1
(14.3)
78.7
(20.8)
120.8
(31.9)
379
(100.1)
419.4
(110.8)
518.2
(136.9)
515.6
(136.2)
E-15
E-22
E-27
E-45
H-60
H-75
H-90
78.7
(20.8)
119.6
(31.6)
140.4
(37.1)
190.0
(50.2)
312.7
(82.6)
367.6
(97.1)
425.5
(112.4)
H-105
L-75
L-90
L-105
466.4
(123.2)
344.9
(91.1)
398.6
(105.3)
445.9
(117.8)
Water flow rate
0-56.8 L/min (0-15 US gallons/min)
Electrical
Input voltage
115/230 V~
230 V~ models are for export only
Line Frequency
50/60 Hz
Rated Current
6.9/3.5 A
Electromagnetic
Compatibility
All Whirlpool models comply with the latest international electromagnetic
compatibility standard (IEC 60601-1-2), as shown below:
Description
Standard
Specifications
Radiated Emissions
CISPR 11
30 MHz
–
1 GHz, Class A
Conducted Emissions, AC Mains
CISPR 11
150 kHz
–
30 MHz, Class A
Harmonic Emissions
IEC 61000-3-2
10 min, Class D (no limits for < 75W)
Flicker
IEC 61000-3-3
120 min (no limits for < 75W)
Electro-Static Discharge (ESD)
IEC 61000-4-2
Contact: ± 2kV, ± 4kV, ± 8kV; Air: ± 2kV, ± 4kV, ± 8kV ± 15kV
Radiated RF EM fields
IEC 61000-4-3
3 V/m, 80% AM at 1 kHz
80MHz
–
2.7GHz
Electrical Fast Transient (EFT)
IEC 61000-4-4
Power Supply Lines, ± 2kV
100 kHz PRF
Surge
IEC 61000-4-5
Line to line ± 0.5kV, ± 1kV
RF Conducted Immunity
IEC 61000-4-6
Power Supply Lines: 3 Vrms, 0.15 - 80MHz; ISM Bands
Input/Output Lines: 3 Vrms, 0.15 - 80 MHz; 6 Vrms, ISM bands, 80% AM at 1 kHz
Magnetic Field
IEC 61000-4-8
30 A
rms
/m, 50Hz or 60Hz
Short Interruption
IEC 61000-4-11
0 % U
T
: 250/300 cycles, 0°, 90°, 270°
Environmental Conditions
Operational
10 - 33
C (50 - 90
F)
Transportation and
Storage
0 - 60
C (32 - 140
F)
Relative humidity
5 - 95%, non-condensing
Atmospheric Pressure
70
– 106 kPa
Thermometer
Range
Dual scale: 5-60
C and 40-140
F
Accuracy
± 1.1°C (2°F)