2. Strip Storage
Storage condition: 4°C~40°C
(39.2°F~10°F), below 85% relative
humidity. Do Not Freeze.
Store your test strips in their ori-
ginal vial only. Do not transfer to
other container.
Store test strip packages in a cool
and dry place. Keep away from
direct sunlight and heat.
After removing a test strip from the
vial, immediately replace the vial
cap and close it tightly.
Touch the test strip with clean and
dry hands.
Use each test strip immediately
after removing it from the vial.
Write the discard date (the date
opened plus 90 days) on the
vial label when you first open it.
Discard remaining test strips 90
days after first opening date.
Do not bend, cut, or alter a test
strip in any way.
Keep the strip vial away from
children since the cap and the test
strip may be a choking hazard. If
swallowed, promptly see a doctor
for help.
3. Control solution storage
Storage condition: Store the control
solution tightly closed at tempera-
tures below 30°C (86°F). Do NOT
refrigerate.
Record the discard date (date ope-
ned plus 90 days) on the control
solution vial. Discard after 90 days.