Error Message
MESSAGE
WHAT DOES
METER SAY?
WHAT IT MEANS
ACTION
Battery is dead.
Please replace.
Appears when the
batteries can not
provide enough
power for a test.
Replace the
batteries
immediately.
Strip has been
used. Please
replace.
Appears when
inserting a used
test strip.
Test with a new
test strip.
The room
temperature is
out of range.
Thus unable to
measure.
Appears
when ambient
temperature is
below system
operation range.
System operation
range is 10~40˚C
(50~104˚F).
Repeat the test
after the meter
and test strip
have reached
the above
temperature.
Appears
when ambient
temperature is
above system
operation range.
Error Message
MESSAGE
WHAT DOES
METER SAY?
WHAT IT MEANS
ACTION
Please ensure
proper operation
then try again.
Remove the strip
after applying
blood to the
absorbent hole.
Re-test with a new
test strip.
Please remove
strip.
Test strip is
inserted after the
meter turns off.
Remove the test
strip.
Meter malfunction.
Please contact
local distributor.
Problem with the
meter.
Review the
instructions and re-
test with a new test
strip. If the above
steps do not work,
please contact the
Customer Service
for help.