7

If the night door is removed during the UV cycle, the function will be cancelled. 
The UV light will switch off, the fan and fluorescent light will switch on and the 
cabinet will return to normal mode ready for use. 

 

UV TIMER ALARM 

The UV timer is pre-set to alarm after a 1000 hrs of tube operation. The tube 
should be replaced at this interval. Although the blue light will still be visible the 
UV power from the tube will be diminished.  
When the tube operation exceeds the pre-set limit, during a UV cycle an audio 
beep will be observed every minute. 
To reset the alarm timer, switch the cabinet off, press and hold the “LIGHT “ 
and the “ UV “ button and switch the cabinet on again. When an audio beep 
sounds release the buttons. The counter will be reset and the system return to 
normal mode.  
 

LOW AIRFLOW ALARM 

The airflow is continuously monitored while the cabinet is running. An audible 
alarm will sound if the airflow drops below a preset level and the red low airflow 
warning light will flash on the control panel. This is an indication that the Pre-
Filter requires replacement – see section 4. If changing the pre-filter does not 
cancel the alarm the cabinet requires re-calibration or a new main HEPA filter. 
 

SYSTEM MONITOR 

The UV timer control circuit has a monitoring component which will indicate a 
system failure. All red warning lights will flash on and off, a continuous audio 
alarm will sound and the fluorescent light will be switched off. 
 
If this occurs the cabinet is unsafe to use. Contact Monmouth Scientific for 
further advice. 
.  

Содержание PCR650

Страница 1: ...ISSUE C 26 th Sept 2008 PCR650 OPERATING AND MAINTENANCE MANUAL ...

Страница 2: ...2 ...

Страница 3: ...LARM 7 SYSTEM MONITOR 7 SECTION 4 8 FILTERS 8 PRE FILTER CHANGING 8 MAIN HEPA FILTER CHANGING 8 SECTION 5 9 MAINTENANCE 9 FUSES 9 LIGHTING 9 UV TUBE 9 AIRFLOW 9 LOW AIRFLOW ALARM 9 CIRCUIT DIAGRAM 10 SECTION 6 11 SERVICING 11 Warning This system must be used in compliance with these instructions and any repairs or maintenance carried out by qualified personnel For parts or service information plea...

Страница 4: ...ean air Laminar Flow over the entire worksurface To further reduce the risk of cross contamination between samples the PCR cabinet is supplied with a time controlled UV light to sterilise the working area between samples 6 individual pre set time functions are provided as standard Circulaire Circulaire Circulaire Circulaire PCR PCR PCR PCR Ca Ca Ca Cabinet binet binet binet External Dimensions 650...

Страница 5: ...e sited in a draught free position ASSEMBLY The cabinet is supplied fully assembled with the main HEPA filter and pre filter fitted TESTING COMMISSIONING A test certificate will be supplied for conformity to CE marking electrical test and airflow ...

Страница 6: ...n of the cycle 1 Press 5 minutes 2 Presses 10 minutes 3 Presses 15 minutes 4 Presses 20 minutes 5 Presses 25 minutes 6 Presses 30 minutes The UV pre set timed cycle will start the fluorescent light will switch off and the UV lamp will switch on The red UV warning light on the membrane panel will flash to indicate the cycle selected 1 flash 5 minutes 2 flashes 10 minutes etc When the UV timing cycl...

Страница 7: ...on again When an audio beep sounds release the buttons The counter will be reset and the system return to normal mode LOW AIRFLOW ALARM The airflow is continuously monitored while the cabinet is running An audible alarm will sound if the airflow drops below a preset level and the red low airflow warning light will flash on the control panel This is an indication that the Pre Filter requires replac...

Страница 8: ...rm is activated Remove the thumbscrews on top of the cabinet and lift off the pre filter retainer Replacement pre filters may be obtained from Monmouth Scientific Part No PF 0023 MAIN HEPA FILTER CHANGING The HEPA filter should be changed by trained personnel Contact Monmouth Scientific for advice ...

Страница 9: ...l panel and may be changed by removing the two securing screws from inside the cabinet and hinging the control panel open The UV light fuse 2A Type F is located on the control circuit board AIRFLOW WARNING This procedure requires adjustment of live circuits and should only be carried out by a qualified person The airflow should be set using a calibrated rotating vane anemometer The fan speed is ad...

Страница 10: ...10 CIRCUIT DIAGRAM ...

Страница 11: ...s Check replace pre filter Check airflow monitor and re calibrate if necessary Check safety interlocks and alarms Inspect electrical components lighting cables etc Issue test report and airflow certificate For parts or service information please contact Monmouth Scientific on 44 0 1278 458090 ...

Страница 12: ...12 Monmouth Scientific Units 5 6 Kilnside East Quay Bridgwater Somerset TA6 4DB Email info monmouthscientific co uk Monmouth Scientific is a trading division of Monmouth Surgical Ltd ...

Отзывы: