background image

 

9

 

SECTION 5 

 

MAINTENANCE 

The cabinet should be isolated from the electricity supply before carrying 
out any maintenance procedures. 
 

FUSES 

Always replace fuses with the correct type and rating: 

 
The main fuses (5A Type T) are located in the mains inlet socket at the back of 
the cabinet. Remove the mains lead and withdraw the fuses using a small 
screwdriver. 
The fan fuse (3.15A Type T) is located on the control circuit board and may be 
changed by removing the two securing screws from inside the cabinet and 
hinging the control panel open. 
 

LIGHTING 

The fluorescent light tube can be accessed from inside the cabinet. The starter 
is located inside the control panel and may be changed by removing the two 
securing screws from inside the cabinet and hinging the control panel open. 
The fluorescent light fuse (2A Type F) is located on the control circuit board. 
 

UV TUBE 

The UV tube can be accessed from inside the cabinet. The starter is located 
inside the control panel and may be changed by removing the two securing 
screws from inside the cabinet and hinging the control panel open. The UV 
light fuse (2A Type F) is located on the control circuit board. 
 

AIRFLOW 

WARNING:

 This procedure requires adjustment of live circuits and should only 

be carried out by a qualified person. 

 
The airflow should be set using a calibrated rotating vane anemometer. The 
fan speed is adjusted by turning the blue potentiometer on the speed control 
circuit board inside the control panel. The speed should be set to achieve a 
downflow of 0.3 – 0.35m/sec. at 100mm above the lower edge of the front fixed 
glazing panel. 
 

LOW AIRFLOW ALARM 

1.  Set the airflow to the correct speed with new pre-filters fitted. 
2.  Switch the cabinet on and place a piece of paper or similar material against 

the pre-filter inlet grille covering all but 14 of the 84 holes to simulate a 
partially blocked filter. 

3.  Unscrew the pressure switch adjusting screw until the low airflow warning 

LED comes on. Screw in until the LED goes out then slowly unscrew until 
the LED comes on again. 

4.  Remove and replace the paper several times to check operation of warning 

LED / audio. 

Содержание PCR650

Страница 1: ...ISSUE C 26 th Sept 2008 PCR650 OPERATING AND MAINTENANCE MANUAL ...

Страница 2: ...2 ...

Страница 3: ...LARM 7 SYSTEM MONITOR 7 SECTION 4 8 FILTERS 8 PRE FILTER CHANGING 8 MAIN HEPA FILTER CHANGING 8 SECTION 5 9 MAINTENANCE 9 FUSES 9 LIGHTING 9 UV TUBE 9 AIRFLOW 9 LOW AIRFLOW ALARM 9 CIRCUIT DIAGRAM 10 SECTION 6 11 SERVICING 11 Warning This system must be used in compliance with these instructions and any repairs or maintenance carried out by qualified personnel For parts or service information plea...

Страница 4: ...ean air Laminar Flow over the entire worksurface To further reduce the risk of cross contamination between samples the PCR cabinet is supplied with a time controlled UV light to sterilise the working area between samples 6 individual pre set time functions are provided as standard Circulaire Circulaire Circulaire Circulaire PCR PCR PCR PCR Ca Ca Ca Cabinet binet binet binet External Dimensions 650...

Страница 5: ...e sited in a draught free position ASSEMBLY The cabinet is supplied fully assembled with the main HEPA filter and pre filter fitted TESTING COMMISSIONING A test certificate will be supplied for conformity to CE marking electrical test and airflow ...

Страница 6: ...n of the cycle 1 Press 5 minutes 2 Presses 10 minutes 3 Presses 15 minutes 4 Presses 20 minutes 5 Presses 25 minutes 6 Presses 30 minutes The UV pre set timed cycle will start the fluorescent light will switch off and the UV lamp will switch on The red UV warning light on the membrane panel will flash to indicate the cycle selected 1 flash 5 minutes 2 flashes 10 minutes etc When the UV timing cycl...

Страница 7: ...on again When an audio beep sounds release the buttons The counter will be reset and the system return to normal mode LOW AIRFLOW ALARM The airflow is continuously monitored while the cabinet is running An audible alarm will sound if the airflow drops below a preset level and the red low airflow warning light will flash on the control panel This is an indication that the Pre Filter requires replac...

Страница 8: ...rm is activated Remove the thumbscrews on top of the cabinet and lift off the pre filter retainer Replacement pre filters may be obtained from Monmouth Scientific Part No PF 0023 MAIN HEPA FILTER CHANGING The HEPA filter should be changed by trained personnel Contact Monmouth Scientific for advice ...

Страница 9: ...l panel and may be changed by removing the two securing screws from inside the cabinet and hinging the control panel open The UV light fuse 2A Type F is located on the control circuit board AIRFLOW WARNING This procedure requires adjustment of live circuits and should only be carried out by a qualified person The airflow should be set using a calibrated rotating vane anemometer The fan speed is ad...

Страница 10: ...10 CIRCUIT DIAGRAM ...

Страница 11: ...s Check replace pre filter Check airflow monitor and re calibrate if necessary Check safety interlocks and alarms Inspect electrical components lighting cables etc Issue test report and airflow certificate For parts or service information please contact Monmouth Scientific on 44 0 1278 458090 ...

Страница 12: ...12 Monmouth Scientific Units 5 6 Kilnside East Quay Bridgwater Somerset TA6 4DB Email info monmouthscientific co uk Monmouth Scientific is a trading division of Monmouth Surgical Ltd ...

Отзывы: