Fig.12
Brake surface
Sand paper
Cord bracket
12
Stop
Tension center
Stop lever
The brake cord should be regularly checked to
ensure that it has not suffered execessive wear. If it
looks worn it should be replaced.
Deposits of dirt on the brake cord and irregularities
on the brake surface may cause the unit to operate
unevenly and will also wear down the brake cord
faster and the unit can become noisy. The brake sur-
face of the flywheel should then be ground off with
a fine sand paper and any dust removed with a clean
dry cloth.
Lift the weight basket to its upper stop position.
Grind with a fine sand paper. See fig. 12.
Grinding is easier to perform if a second individual
cautiously and slowly pedals the cycle.
NOTE: Always keep the brake belt contact surface
clean and dry. No lubricant is allowed to be used.
We recommend to replace the brake belt when
cleaning the contact surface.
As regards assembly and adjustment of the brake
belt, see previous page.
Return strap
Hole fore cord knot
Fig.11
FLYWHEEL BRAKE SURFACE - BRAKE BELT