Chest belt ANT+ comes with as standard.
In order for the heart rate measurement to work, it is
necessary that the pulse belt is correctly positioned, see
Fig: Placement chest belt.
Moisten the electrodes before use,
see
Fig: Moistening the electrodes.
To make contact with the HR receiver on the bike, the
distance should not be more than 100 cm. At the first
usage it is important to first pair the chest belt with the
display by standing close to it so the display can receive
the Control signal (max. 60 cm).
HEART RATE
Fig: Placement chest belt
Fig: Moistening the electrodes (1)
PULSE STANDARDS
The following pulse standards can be used:
•
Standard uncoded 5K belts (5-5.6 kHz)
•
Chest belts with only ANT+
Fig: Zero position (1)
SCALE ADJUSTMENT
NOTE!
It is important that you are
very precise when you
adjust or reset the scale.
Fig: Scale adjustment
1) Locking screw
Turn on the power to the ergometer. Release any tension
in the brake belt by lifting the pendulum / pointer to
slightly over 4 kp. Hold the pendulum in this position for
a few seconds, then move it down again. Check that the
pointer conforms to the zero index.
If the pointer is not in zero, the scale must be adjusted.
First, loosen the locking screw and then change the
position of the scale so that the 0-index is in line with the
pointer. Tighten the screw after adjustment. See
Fig: Scale
adjustments.