20
!
VII. Operating Procedures
(continued)
The message periodically appears at the top of the LCD screen and means
the
Monarch
is in an Off Cycle and that it is not currently providing electrical stimulation.
This is normal and indicates the
Monarch
is functioning properly. During this Off Cycle
the current amplitude cannot be increased, but it can be decreased.
If you wake up and the
Monarch
is powered off, first make sure all the connections are intact.
Then, press the ON/OFF key. If the
Monarch
does not turn on, remove the battery and insert
a fully-charged battery. Press the ON/OFF key again. If there are recurrent device
turn-offs, contact your physician.
Be sure to always begin treatment with a fully-charged battery and to lock the
Monarch
after adjusting the settings.
After Treatment
When treatment is complete, unlock the
Monarch
by pressing the LOCK key for two seconds.
Turn the
Monarch
off by pressing ON/OFF key.
Unplug the lead wire from the
Monarch
.
Unplug the lead wire from the electric patch wire.
Gently remove the electric patch and dispose. Do not reuse.
Gently wash or clean skin of the forehead.