StakMax
®
Plate Handling System User Guide
0112-0133 Rev B
107
•
If you want to save the data from each plate or set of plates to a
different data file, select
Open a New Protocol
. For example, if
you have 50 sample plates and the assigned protocol has one
Plate section, the software opens 49 copies of the protocol and
saves the data to 50 data files. Alternatively, if you have 4 sets of 5
plates (1 control plate and 4 sample plates) and the assigned
protocol has 5 plate sections, the software opens 4 instances of
the protocol for each set of plates and saves the data to 4 data
files. This option allows you to easily automate the processing of
plate sets without the need to develop a script.
•
If you want to save the data from all of the plates to the same data
file, select
Append a Plate Section
. For example, if you have 50
sample plates and the assigned protocol has one Plate section,
the software adds 49 Plate sections to the original protocol and
the data from all of the plates is saved in a single data file.
Source Protocol:
Click Assign to select a protocol. The Choose a
Protocol for reading window appears. See
.
Figure 5-20
The Choose a Protocol for reading window.
•
Select a protocol and then click
Open
.
Содержание StakMax
Страница 1: ...StakMax Plate Handling System User Guide 0112 0133 Rev B May 2010...
Страница 6: ...Contents 6 0112 0133 Rev B General Specifications 141 Diagrams 142...
Страница 10: ...Safety 10 0112 0133 Rev B...
Страница 14: ...Introduction 14 0112 0133 Rev B...
Страница 59: ...StakMax Plate Handling System User Guide 0112 0133 Rev B 59 Figure 3 9 Pressing the Power Reset button...
Страница 80: ...Installation and Setup of SpectraMax L instrument 80 0112 0133 Rev B...
Страница 116: ...Operation 116 0112 0133 Rev B...
Страница 146: ...System Diagrams and Specifications 146 0112 0133 Rev B...