SpectraMax M2 and M2e Microplate Reader User Guide
40
0112-0102 E
Note:
If an error occurs while the plate drawer is closed, and you need to remove a
plate, see
Opening the Drawer Manually on page 43
.
Error
Code
Error Message
Notes
100
command not
found
Command string not recognized.
101
invalid argument
Command argument not recognized.
102
too many
arguments
Too many arguments after command.
103
not enough
arguments
Missing arguments.
104
input line too long
Too many characters in the input line.
105
command invalid,
system busy
Instrument could not perform the give command because it was busy
doing another task.
106
command invalid,
measurement in
progress
Instrument could not perform command because a measurement was
in progress.
107
no data to transfer
Inputting transfer when there is no data in the buffer.
108
data buffer full
Too many data sets in the buffer. Can be caused by setting up a long
kinetic read and then disconnecting computer or the SoftMax Pro
Software is preempted by another application.
109
error buffer
overflow
More than 65 errors in the buffer, clear the buffer.
110
stray light cuvette,
door open?
Cuvette door open while doing a read.
111
invalid read
settings
Table 7-2: Error Codes 100-199: Unrecognized Command Errors Sent From the Computer